FMX 320/FMX 640 SERIES IR CAMERA USER MANUAL
22
11-1 Camera(s) not showing, camera(s) lagging, or software crashing
•
Close and reopen software
•
Reconnect power & Ethernet cables
•
Ensure that camera has been powered on for a few minutes
•
Verify that static IP address is correct:
◦
Default IPv4 of 192.168.10.110
◦
Default subnet mask of 255.255.255.0
•
Restart computer
•
Verify that correct software is installed
•
Uninstall & reinstall software, running as administrator
•
Verify that firewall is not blocking software
•
Try newer or different version of IR Flash STM software
11-2 Camera out of focus
•
For FMX 320 and 640 cameras:
◦
Adjust focus ring
◦
Close and reopen software
◦
Try newer or different version of IR Flash STM software
◦
Restart computer
11-3 Temperature readings are incorrect or facial recognition is suboptimal
•
Close and reopen software
•
Check that software settings are correct:
◦
Fixed temperature of 35 °C or 37 °C, depending on readout of Temperature
Reference Source
◦
Temperature Reference Source has crosshair over it
•
Remove masks & glasses
•
Camera is proper distance from person
•
Camera is at proper height
•
Temperature Reference Source in view, perpendicular to camera, but not blocked
•
Person is looking directly at camera lens, not at an angle
•
Camera is in focus
11. Troubleshooting