FM PLUS P SERIES IR CAMERA USER MANUAL
19
□
Height of camera
□
Average male 70 ± 4 inches; Average female 65 ± 4 inches
□
5’4” – 5’8” will image 4’8” to 6’4” individual
□
Larger will require a height adjustment to camera
□
Distance of camera to individual being screened is: 8 – 10 feet
□
The individual’s face and temperature reference source should be in the frame
□
Distance to temperature reference source: 12 feet
□
Behind the individual being screened yet visible within the frame of the image
□
Protect the camera from exposure to direct sunlight to avoid damage to the camera
sensor
10-6 Temperature reference source
□
On tripod or other stable stand, in front of background, and positioned for visibility
within the frame of the image
□
Temperature is set at 35 °C (95 °F) or 37 °C (98.6 °F)
□
Warm-up time is 15 minutes
10-7 Software
□
Set up your software according to the instructions provided
□
Documentation can be found online on the IR Flash STM web page under the Downloads
section or at this address:
https://infraredcameras.com/ir-flash-stm-manual/
10-8 Final Step
Final validation is dependent upon a signed copy of this checklist from each person
responsible for each site.
Upon completion please sign below:
ICI representative:
Company representative:
Signature: _____________________________ Signature: _______________________________
Print Name: ____________________________ Print Name: ______________________________
Date: ________________