IceCure Medical Ltd.
DSR3210000 rev. D
ProSense™
Confidential
User Manual
17
Return to first page
2.2.7
Proper use
Caution
Do not disconnect the probe from the system during FREEZE, EXTRACTION,
or THAW unless you are required to do so.
Caution
Do not use the workstation or the liquid nitrogen Dewar for any purpose
other than
operating the ProSense™ cryoablation system.
Caution
The liquid nitrogen Dewar s
upplied with the ProSense™ cryoablation system
is a dedicated system part and should not be used for any other purpose.
Make sure to use only IceCure approved and labeled Dewars for your
system.
Caution
Dewars should always be stored with their lids in place.
Caution
Only Dewars and lids supplied by IceCure Medical may be used with the
ProSense™ cryoablation system. Make sure to use only an approved IceCure
Dewar for your system.
Warning
Always turn OFF the cryoablation system and lock the wheels when not in
use.
Before you start a procedure, lock at least two of the four wheels.
Caution
The ProSense™ cryoablation system should be moved with care i
n order to
avoid damage to the system or other clinical equipment.
Warning
After extraction step, before removing the cryoprobe from the tissue, make
sure that the freeze effect has been deactivated so that the probe can easily
be removed from the tissue.