IceCure Medical Ltd.
DSR3000000 rev.L
Confidential
User Manual
82
IceSense3™
Warning
In case of software crash, switch OFF the mechanical ON/OFF button and unplug the electrical cable.
Call IceCure Medical for technical service before restarting the IceSense3™
cryoablation system.
5.5.2.1
Emergency Stop button
The fastest way to end a procedure is to press the Warm button in the handle or on the screen.
However, in case of emergency, you may press the RED ROUND Emergency Stop button located on
the right side of the main chassis. The Emergency Stop button is designed for emergency shutdown
of the unit. Pressing this button immediately turns off the system.
Switch OFF the mechanical ON/OFF button and unplug the electrical cable.
Figure 71: The Emergency Stop button located on the main chassis (green arrow)
Warning
If a liquid nitrogen leak is detected at any time, PUSH the Emergency Stop button immediately.
Caution
Push the Emergency Stop button ONLY when there is NO other choice. Wherever possible, the
system should be shut down by the standard procedure. Excessive use of the Emergency Stop
button may damage the system.
After pressing the emergency stop button, wait for "passive thaw" before extracting the cryoprobe
from the tissue.