PARTS OF THE TONOMETER
CLEANING THE PROBE BASE
The probe base can be reused after careful cleaning. Make sure
that the probe base is completely dry before using it.
1.
Unscrew the probe base collar and carefully remove the probe
base from the tonometer.
2. Invert the probe base over a container filled with 70-100%
isopropyl alcohol solution, allowing the probe base to drop into
the container and let soak for 5 - 30 minutes.
3. Dry the probe base by blowing some compressed air into it.
4. Carefully reinsert the completely dry probe base, screw the
collar back into the tonometer.
Please refer to the manual for more detailed instructions of
changing and cleaning the probe base.
CHANGE/CLEAN THE PROBE BASE REGULARLY TO AVOID
UNNECESSARY SERVICE.
www.icaretonometer.com
Display
Navigation buttons:
Left, Down, Up & Right
Main button
Forehead support adjusting wheel
Forehead support
Collar
Probe base
Please refer to the manual
for additional information.