Icare CLINIC, EXPORT and PATIENT Instruction Manual for Patients   |   

ENGLISH

 

4.4.2

 

Managing IOP results and creating reports 

IOP measurements are listed under the 

Measurements

 tab. OD means the right eye. OS means the 

left eye. The listed measurements can be filtered either by entering 

From

 and 

To

 dates or by 

selecting a HOME use period from the 

Select period

 drop down list. 

The quality of the measurement, as calculated by the tonometer, is shown to the right of the 

displayed IOP value. The term 

Excellent

 for quality indicates a low variation among the IOP data 

sampled by the tonometer during a measurement. The term 

Good

 indicates fairly low variation 

among the data sampled. 

Poor

 indicates higher but still acceptable variation of the data sample for 

IOP values 19 mmHg or lower. 

Rejected

 indicates high variation in the sampled data and the IOP 

result should not be used for clinical determination.

 

  

Содержание EXPORT TS03

Страница 1: ...Icare CLINIC Icare EXPORT Icare PATIENT app INSTRUCTION MANUAL F O R P A T I E N T S ENGLISH...

Страница 2: ...pe TS04 Instruction Manual for Patients TS02 014 EN 1 4 The information in this document is subject to change without prior notice In a conflict situation the English version prevails Copyright 2019 I...

Страница 3: ...on of the Icare PATIENT app 9 4 2 Start up and shutdown 10 4 3 Using the Icare HOME tonometer Icare CLINIC Icare EXPORT and the Icare PATIENT app together 11 4 4 Using Icare CLINIC 13 4 4 1 Introducti...

Страница 4: ...g the Icare EXPORT or PATIENT application is set to the correct time and date PRECAUTION Precaution Always consult your healthcare provider before changing your treatment 2 INTRODUCTION This is the pa...

Страница 5: ...r Icare account with information on how to access the service You need to use either Icare EXPORT on a computer or the Icare PATIENT app on a smartphone or a tablet to upload IOP data to the Icare CLI...

Страница 6: ...colors DirectX 9 compatible Graphics Card NET Framework 4 6 1 or greater Operating System Windows 7 Windows 8 or Windows 10 Internet connection when used to upload results to Icare CLINIC To install I...

Страница 7: ...terms appears Confirm the terms by ticking I agree and click Next 3 A window for selecting the installation folder appears Select a folder to install the application Select Everyone to allow Icare EXP...

Страница 8: ...oud mode Icare EXPORT sends the measurement results to your healthcare provider using the Icare CLINIC service Local mode should be selected to manage the IOP measurement results locally in the comput...

Страница 9: ...e PC 4 1 3 System requirements and installation of the Icare PATIENT app Minimum system requirements for the Icare PATIENT app Android smartphone or tablet with Operating system v 4 0 1 or later with...

Страница 10: ...re PATIENT app in the Google Play application store and then follow the installation instructions displayed on the smartphone or tablet screen 4 2 Start up and shutdown To start Icare CLINIC open a we...

Страница 11: ...To start the Icare PATIENT app connect an Icare HOME tonometer to your smartphone or tablet using the USB OTG cable To shut down the Icare PATIENT app use the application management functionality of y...

Страница 12: ...installed The internal clock gets automatically updated to the time of the connected computer smartphone or tablet PRECAUTION Ensure that the computer smartphone or tablet running the Icare EXPORT or...

Страница 13: ...personal information This information includes your personal forehead and cheek support settings You should check that these settings are the same as documented in the patient profile before self mea...

Страница 14: ...LISH Measurement list shows your measurements either your self measured readings indicated by the device type Icare HOME or those measured by your healthcare provider By default the IOP measurements f...

Страница 15: ...m the Select period drop down list The quality of the measurement as calculated by the tonometer is shown to the right of the displayed IOP value The term Excellent for quality indicates a low variati...

Страница 16: ...sured on the y axis The measurement time and the calculated IOP for individual measurements are revealed by moving the cursor with the PC mouse over the IOP results on the graph The user can zoom the...

Страница 17: ...www icarefinland com 17 Icare CLINIC EXPORT and PATIENT Instruction Manual for Patients ENGLISH Chart type Scatter The chart type Scatter displays the data points without connecting them with lines...

Страница 18: ...Patients ENGLISH The chart type Diurnal positions all the IOP data points to a 24 hour x axis Thus all data points measured at the same time of the day will show at the same point on the x axis The I...

Страница 19: ...ng the PRINT button on the report Click the SHOW TABLE button to open and display the measurement list below the graph 4 5 Using Icare EXPORT to upload IOP results from the Icare HOME tonometer Once y...

Страница 20: ...IC database are removed from the memory of the Icare HOME tonometer Click the LOG IN link on the upper right hand side to login to Icare CLINIC if enabled by your healthcare provider for viewing your...

Страница 21: ...you have made some measurements perform the following steps 1 Connect the Icare HOME tonometer to the smartphone or tablet using the USB OTG cable NOTE The Host plug of the cable must be connected to...

Страница 22: ...ed to the smartphone or tablet Disconnect and reconnect the cable to the smartphone or tablet and or restart the Icare PATIENT app You have uploaded the IOP measurement result to Icare CLINIC but they...

Страница 23: ...ometers are always displayed as integers of mmHg For the proper use of the Icare EXPORT on a PC the user s network environment must allow Icare EXPORT to access the internet using TCP port 443 Failure...

Отзывы: