15
AfiMilk MPC
Installation Manual
7.3.1.2
Water test adjustment and setting reference
7.3.1.2.1
Global Test description
o
Perform one (or direct two consecutive) water tests procedures as described above
(same amount of water) for each milk meter in the parlor.
o
Check if on or more meters deviate in result against the other meters (result
difference > 0,25 kg on average difference of all meters). If so, re-check this meter(s)
and approve result or perform a re-test.
Example form to store and analyse the collected data
In this example, two measurements are done. Meter no. 4 shows a bigger difference as
all other meters. This meter is re-checked and considered to be Ok. This value is
accepted.
o
If within the milk meter the two water tests are differ no more then 0,1 kg, that
meter is Ok. If the difference is more then 0,1 kg a third measurement should be
performed. The three measurements should differ no more then 0,2 kg. If so, re-test
and replace one or more existing measurements.
Conduc-
Conduc-
Meter G-value
Display
Unster
Diff.
tivity
afwijking
Display Unster
Diff.
tivity
1
53
12,20
11,44
0,76
11,7
12,32 11,48
0,84
11,8
2
53
12,30
11,56
0,74
11,6
12,26 11,46
0,80
11,8
3
53
12,25
11,55
0,70
11,6
12,20 11,48
0,72
11,7
4
53
12,28
11,20
1,08
11,6
12,32 11,20
1,12
11,6
5
53
12,30
11,48
0,82
11,5
12,40 11,60
0,80
11,6
6
53
12,44
11,68
0,76
11,8
12,30 11,48
0,82
11,9
7
53
12,18
11,26
0,92
11,7
12,46 11,66
0,80
11,8
8
53
12,20
11,40
0,80
11,7
12,28 11,46
0,82
11,7
9
53
12,24
11,46
0,78
11,6
12,26 11,50
0,76
11,6
10
53
12,16
11,36
0,80
11,5
12,20 11,34
0,86
11,6
0,82
Avarage water test diff.
Pre watertest
2e water test
Vacuum level (kPa)
42,0
Parlor: Harringbone (2x5)
Date: 30-7-2018
Test person Joe
three within 0,2 ( else replace faulty measurements)
two measurements with mutual difference max. 0,1 kg
Afimilk MPC Setting reference values by water test
Farm:
1265897
Technical check
Farm tesing
Hygienic check
Cow road
Performer:
Cow city