21
Frequency
Number of cycles
Maintenance operation
Monthly
50
Clean the door seal
Clean the filter inside the chamber and in the clean
water tank
Clean the chamber the trays and the rack
Clean the external surface
Every 3 months
200
Clean the distilled water tank
Every year
800
Replace the door seal
Select Settings from the main menu and tap “Maintenance”, to enter in the menu. Select the
item to check the information.
Clean the distilled water tank
Disconnect the main cable.
Drain the tank completely using the drain connector at the back of the sterilizer and leave it connected
into the connector in an open position.
Clean the internal surface with a soft sponge and a small soft brush for the areas difficult to reach using
and a distilled water.
Remove the filter and clean it with a small soft brush and mild soap, rinse it with distilled water, and put it
back in to the position.
Replacement of the bacteriological filter.
The bacteriological filter is in the back of the sterilizer. Unscrew
the filter by hand anti-clockwise.
Place the new bacteriological filter.
Screw the new filter by hand clockwise.
Note: Do not operate sterilizer without filters in place.
Clean chamber, door seal ring, trays, and tray rack.
Remove the trays and tray rack from the chamber.
Clean trays, rack and inside of the chamber with mild soap.
Rinse the trays, rack and inside of the chamber with a smooth cloth and distilled water.
Examine door seal for possible damage.
Clean door seal and mating surfaces with a damp cloth.
Note: Do not use bleaching agents or any abrasive materials or substances in the chamber. Failure to
comply may cause damage to the chamber and/or other components.
Caution: To prevent burns, let the unit cool before cleaning gaskets and touch the surface.