background image

Instructions Manual

For Models: STE-18-T

STE-23-T

Содержание STE-18-T

Страница 1: ...Instructions Manual For Models STE 18 T STE 23 T...

Страница 2: ...ully and follow all installation instructions Ningbo Ican Machines Co Ltd No 77 Yunlin East Road Gulin Town Ningbo China www icanclave com European Representative Icanclave Europe S L Juan Ramon Jimen...

Страница 3: ...mber 14 Select the program 14 Start the sterilization program 15 End of cycle 15 Manually interruption of the cycle 16 Delay start cycle 17 Test Program 18 Descale 19 Data 19 Print report 20 Print Lab...

Страница 4: ...nnect only to a properly grounded outlet Do not cover or block any openings on this appliance Use this appliance only for its intended use as described in this manual Do not exceed the maximum weight...

Страница 5: ...Tap to open the door The door is open Fill the tank if it is flashing From an external container Tap until filling the tanks appears Drain the tank if it is flashing The capacity is 1 5 liters 1 Repor...

Страница 6: ...Door seal ring 12 USB port 20 Safety valve 5 Door 13 Distilled water tank Inn 21 Valves vent 6 Door switch 14 Used water tank 22 Used water tank Vent 7 Distilled water tank 15 Condenser vent 8 Touch s...

Страница 7: ...rack 1 3 Draining hose 2 4 Door Seal 1 5 Try handle 1 6 Instructions Manual 1 7 Calibration certificate 1 3 Installation General conditions Position the device on a plane surface with a minimum capac...

Страница 8: ...tch On and open the door to remove all of the inner contents for unpacking Connect the power cord to an outlet of the appropriate voltage Turn on the main power switch on the right side After switchin...

Страница 9: ...left Attention The capacity of the used water tank is approximately 2 liters Preparation of sterilization materials For the most effective sterilization and to preserve the sample please follow below...

Страница 10: ...he screen will show the current configuration Tap to enter the menu Touch the parameter that needs to be change it will light orange by pressing or by using the numbers adjust it To save press Languag...

Страница 11: ...o save and back Pre Heating Tap Pre heating button The screen will show the current configuration Tap to enter the menu Tap ON or OFF to select tap to save and back Fill Select Fill Tap to enter the m...

Страница 12: ...t the cycle name for example Quick The screen will show the current configuration Tap to enter the menu Touch the parameter or by pressing select the number that needs to be changed it will light red...

Страница 13: ...tings Tap Click here to reset all settings The screen will show confirm to reset all settings To confirm press To deny press Maintenance Tap Settings then Maintenance You can see the information of ho...

Страница 14: ...elect Programs Tap the program to select in the screen will appear the information of selected program as the pressure temperature time this may vary depending on the environmental conditions and the...

Страница 15: ...appear next to the door status icon Caution Always use the tray handle to load or unload the tray into the sterilizer Failure to do so can result in burning If the label printer is connected At the e...

Страница 16: ...ide the sterilizer must be considered not sterile Caution Depending on the phase of the cycle steam and water can escape from the sterilization chamber when you open the door Delay start cycle Tap Del...

Страница 17: ...Helix test into the chamber then close the door and Tap After finishing the cycle you may check the indicator and evaluate the result according with the instructions of the test manufacturer B D test...

Страница 18: ...e difference between the max Temperature and the Min is above 3 C it will show void That means the result of the test is fail You need run the vacuum test again after the chamber has cooled down Desca...

Страница 19: ...Min HoldingTime 4 0Min Time Temp Pressure Start 12 28 17 089 0C T1 12 31 32 087 1C 075 0kPa T2 12 33 43 110 2C 052 0kPa T3 12 36 37 088 9C 075 0kPa T4 12 39 20 114 7C 053 7kPa T5 12 43 37 087 9C 075 0...

Страница 20: ...the serial number of the steam sterilizer is E00001 and the cycle number is 0012 The file name of the file will be 01001200 log The first two numbers represent machine number The middle four numbers...

Страница 21: ...clean it with a small soft brush and mild soap rinse it with distilled water and put it back in to the position Replacement of the bacteriological filter The bacteriological filter is in the back of t...

Страница 22: ...ve the door seal ring carefully by hand Clean the door seal ring carefully with a smooth cloth with distilled water Moisten the new seal with distilled water Insert the new seal and tap in sequence as...

Страница 23: ...chamber Power off run a new cycle Contact your supplier if error persists E13 Vacuum failed Power off run a new cycle Contact your supplier if error persists E20 Program manually interrupted Reset th...

Страница 24: ...or Action Signal of the wrong position of the door 5 Thermostat on chamber heating resistors Protection for possible overheating of the chamber heating resistors Action Interruption of the power suppl...

Страница 25: ...icium oxide sio2 1mg I 1 0mg kg Iron 0 2mg I 0 1mg kg Cadmiun 0 005mg I 0 05mg kg Lead 0 05mg I 0 1mg kg Rest of heavy metals 0 1mg I 0 1mg kg Chloride 2mg I 0 1mg kg Phosphates 0 5 mg I 0 1mg kg Cond...

Страница 26: ...Quick S 134 210 4 11 15 Unwrapped solid material 1 00 0 8 Standard B 134 210 4 30 49 Unwrapped solid material 4 50 1 50 Single wrapped solid or hollow material 4 00 1 20 Soft B 121 110 20 38 54 Unwra...

Страница 27: ...Quick S 134 210 4 15 20 Unwrapped solid material 1 20 0 8 Standard B 134 210 4 38 57 Unwrapped solid material 5 00 1 50 Single wrapped solid or hollow material 4 50 1 20 Soft B 121 110 20 46 62 Unwra...

Отзывы: