background image

 

 
Last update: 

2022-06-13 

 

 IBS 

Precision Engineering • Eindhoven • NL • +31 40 290 1270

 

User Manual |

 

P3

 

 

1

 

Approvals and safety considerations 

The ECL202/ECL202e is compliant with the following CE directives: 

Safety:

 61010-1:2001 

EMC:

 61326-1, 61326-2-3 

To maintain compliance with these standards, the following operating conditions must be maintained: 

 

All I/O connecting cables must be less than three meters in length 

 

AC power cables must be rated at a minimum of 250 V and 5 A 

 

AC power must be connected to a grounded mains outlet rated less than 20 A 

 

Power supply must have CE certification and provide safety isolation from the mains according to IEC60950 
or 61010. 

 

Sensors must not be attached to parts operating at hazardous voltages in excess of 30 VRMS or 60 VDC 

 

All external connections must be SELV (Safety Extra Low Voltage). 

Use of the equipment in any other manner may impair the safety and EMI protections of the equipment. 

 

2

 

Helpful Technical Support Documents Online 

The IBS Precision Engineering’s website has a large selection of technical documents (TechNotes and Application 

Notes) in the Technical Library. These documents provide detailed descriptions of the operation and use of the 
products of IBS Precision Engineering. 

The Technical Library can be accessed at: 

https://www.ibspe.com/expertise/technical-resources

  

Some of the titles include: 

 

Understanding Capacitive and Inductive Sensors 

 

Comparing Capacitive and Inductive Sensors 

 

Z-height Measurement with Non-contact Sensors 

 

Sensor Operation and Optimization 

 

Using Capacitive Sensors in Vacuum Applications 

 

Understanding Electrical Runout When Using an Eddy-Current Sensor for Roundness Measurements 

 

Inductive Probe Cabling Considerations 

 

 

Содержание ECL202

Страница 1: ...www ibspe com User Manual P1 ECL202 ECL202e ...

Страница 2: ...RS 5 5 1 LED RANGE INDICATOR 5 5 2 RESETTING OFFSET 5 5 3 SETPOINT BUTTON 5 6 ANALOGUE OUTPUT SIGNAL 6 6 1 INTERPRETING THE OUTPUT VOLTAGE 6 6 2 CHANGE IN GAP CALCULATION 6 7 REMOTE OFFSET AND SETPOINT 6 8 SETPOINT SWITCH OUTPUT 6 9 BANDWIDTH SELECTION ECL202 ONLY 6 10 MAXIMISING PERFORMANCE 7 10 1 EXTENSION CABLES 7 10 2 PROBE MOUNTING 7 10 3 UNGROUNDED TARGETS 7 10 4 MULTIPLE SENSORS 7 11 CONNEC...

Страница 3: ... or 60 VDC All external connections must be SELV Safety Extra Low Voltage Use of the equipment in any other manner may impair the safety and EMI protections of the equipment 2 Helpful Technical Support Documents Online The IBS Precision Engineering s website has a large selection of technical documents TechNotes and Application Notes in the Technical Library These documents provide detailed descri...

Страница 4: ...consists of driver electronics and a probe calibrated for a specific material and range The calibration information is detailed on a calibration certificate which is shipped with the system The ECL202 provides a linear analog voltage proportional to changes in the target position and a digital switched setpoint output with a user programmed switching setpoint 4 Quick start instructions 1 Connect t...

Страница 5: ...not a reliable indication of the target position This function establishes a repeatable master point for reference measurements 1 Place good part in the measurement area 2 Position probe to center 20 of range center indicator LED 3 Press Offset button 4 All subsequent measurements indicate deviation from center of range 5 V 5 2 Resetting Offset Hold the Offset button for four seconds to remove any...

Страница 6: ... buttons can be activated remotely Each remote input connects to an opto isolator The functions are activated by applying 15 24 V to the remote control input terminals Note Because the remote operation mimics front panel operation activating the Offset function for more than four seconds will restore factory default value for Offset 8 Setpoint switch output When the output voltage is more positive...

Страница 7: ...asured Otherwise custom calibration will be required 10 3 Ungrounded Targets Ungrounded targets have the potential to inject noise into the sensor If the output is unusually noisy be sure the target is grounded On moving rotating targets this can be accomplished with a small metal brush or thin piece of metal which is connected to ground 10 4 Multiple Sensors When multiple sensors are used with th...

Страница 8: ...t function 15 24 VDC Activates the function Remote Threshold Opto isolator input for Setpoint function Switch Contact 1 of switched output Open 30 VAC 60 VDC max Closed 250 mA max Switch Contact 2 of switched output Bottom Front Connector Power and Analog Output Power In Input Power Internally connected together Power Ground Power Ground Output Signal Ground Reference for output signal voltage Out...

Страница 9: ...log Output Update Rate 15 μS Setpoint Switch Output Solid state switch closure On state 2 5 Ω 250 mA max Off state 30 VAC 60 VDC max Remote Setpoint and Offset Inputs 15 24 VDC to activate 3 7 mA Opto isolator inputs Driver Operating Environment 4 C 50 C IP40 Probe Operating Environment Standard Probes 25 C to 125 C IP67 High Temp Probes 25 C to 200 C IP63 Actual values depend on probe and range a...

Страница 10: ... C U5 Probe 0 1 F S C U8 Probe 0 04 F S C U12 Probe U18 Probe U25 Probe U38 Probe U50 Probe Temperature Coefficient Probe Nonferrous U3 Probe 0 04 F S C Over 15 C to 65 C Temperature range U5 Probe 0 04 F S C U8 Probe 0 02 F S C U12 Probe 0 02 F S C U18 Probe 0 01 F S C U25 Probe U38 Probe U50 Probe Ferrous U3 Probe 0 04 F S C 15 C to 35 C 0 08 F S C U5 Probe 0 01 F S C U8 Probe 0 04 F S C U12 Pro...

Страница 11: ...Last update 2022 06 13 IBS Precision Engineering Eindhoven NL 31 40 290 1270 User Manual P11 13 Mechanical detail ...

Отзывы: