Analogue output signals 

Each channel’s output sign

al is a ±5 VDC analog voltage. The output voltage is proportional to the probe-target 

gap. As the probe-target gap increases, the voltage becomes more positive. See the included calibration 
certificate for specific information. 

Interpreting the Output Voltage 

Output voltage change for a given change in the probe-target gap is called sensitivity. The sensitivity of the 
sensor is listed on the calibration certificate. 

Change in gap calculation: 

Gap Change = Voltage Change / Sensitivity 

For example: With a sen

sitivity of 1V/2 μm and a voltage change of +3 V, the probe

-target gap has increased by 

6 μm.

 

 

7

 

Probe mounting 

When multiple probes are mounted together, they must be separated by at least 3 probe diameters to prevent 
interference between the channels. The area within 3 probe diameters to the sides and 1.5 diameters behind 
should be kept clear of any metallic objects other than the object to be measured. Otherwise, custom calibration 
will be required. If this is not possible, field calibration may be required. 

 

 

 

8

 

Extension cables 

Sensors which are calibrated with a probe extension cable must be operated with the extension cable installed to 
meet specifications. Operating without the extension cable will result in inaccurate measurements. 

 

Содержание ECL150

Страница 1: ...www ibspe com User Manual P1 ECL150 ...

Страница 2: ... 3 1 DESCRIPTION 4 4 QUICK START INSTRUCTIONS 4 5 LED RANGE INDICATOR 4 6 ANALOGUE OUTPUT SIGNALS 5 7 PROBE MOUNTING 5 8 EXTENSION CABLES 5 9 UNGROUNDED TARGETS 6 10 MULTIPLE SYSTEMS 6 11 SETTING BANDWIDTH 6 12 CONNECTING TO THE ECL150 7 12 1 SYNC CONNECTIONS 7 13 OUTPUT 1 4 CONNECTIONS 8 14 OUTPUT 5 8 CONNECTIONS 8 15 SPECIFICATIONS 9 16 MECHANICAL DETAIL 11 17 POWER SUPPLY 11 ...

Страница 3: ...ating at hazardous voltages in excess of 30 VRMS or 60 VDC All external connections must be SELV Safety Extra Low Voltage Use of the equipment in any other manner may impair the safety and EMI protections of the equipment 2 Helpful Technical Support Documents Online The IBS Precision Engineering s website has a large selection of technical documents TechNotes and Application Notes in the Technical...

Страница 4: ...rovides linear 5 V outputs for each channel more positive output voltage indicates the probe is further from the target The calibration information is detailed on a calibration certificate which is shipped with the system 4 Quick start instructions 1 Connect the probes to the matching channels on the ECL150 NOTE Each channel is calibrated to a specific probe identified by a channel number marker o...

Страница 5: ...hange Voltage Change Sensitivity For example With a sensitivity of 1V 2 μm and a voltage change of 3 V the probe target gap has increased by 6 μm 7 Probe mounting When multiple probes are mounted together they must be separated by at least 3 probe diameters to prevent interference between the channels The area within 3 probe diameters to the sides and 1 5 diameters behind should be kept clear of a...

Страница 6: ...used on the same target When multiple ECL150s are used with the same target the systems must be synchronized for best performance An interconnecting cable connects between the Sync In and Sync Out connectors Single systems do not come with these connectors installed 11 Setting bandwidth The system is shipped with the bandwidth setting listed on the calibration certificate based on order specificat...

Страница 7: ...tor 4 Pin M12 Male Power Cable PN 6203 0400 Pin Wire Color Connection 1 Brown 12 to 24 VDC Internally Connected 2 White 12 to 24 VDC 3 Blue Power Ground Internally Connected 4 Black Power Ground 12 1 Sync Connections Sync connectors are only installed when multiple systems are sold together Connect a sync cable Lion Precision Part Number 6203 0820 from the SYNC OUT of one system to the SYNC IN of ...

Страница 8: ...ection 1 White Ch1 Out 2 Brown Ground 3 Green Ch2 Out 4 Yellow Ground 5 grey Ch3 Out 6 Pink Ground 7 Blue Ch4 Out 8 Red Ground 14 Output 5 8 Connections Device Connector 8 Pin M12 Female Output Cable PN 6203 0810 Pin Wire Color Connection 1 White Ch5 Out 2 Brown Ground 3 Green Ch6 Out 4 Yellow Ground 5 Grey Ch7 Out 6 Pink Ground 7 Blue Ch8 Out 8 Red Ground ...

Страница 9: ...tput Update Rate 15 μS Driver Operating Environment 4 C 50 C IP64 Probe Operating Environment Standard Probes 25 C to 125 C IP67 High Temp Probes 25 C to 200 C IP63 Actual values depend on probe and range and are listed on the calibration certificate shipped with the product Contact Lion Precision for replacement certificates In High EMI environments 10 V m output noise levels may rise to 60 mV 0 ...

Страница 10: ...s U3 Probe 0 08 F S C U5 Probe 0 10 F S C U8 Probe 0 04 F S C U12 Probe 0 1 F S C U18 Probe U25 Probe U38 Probe U50 Probe Temperature Coefficient Probe Nonferrous U3 Probe 0 04 F S C Over 15 C to 65 C Temperature range Except where noted U5 Probe 0 04 F S C U8 Probe 0 02 F S C U12 Probe 0 02 F S C U18 Probe 0 01 F S C U25 Probe U38 Probe U50 Probe Ferrous U3 Probe 0 08 F S C U5 Probe 0 10 F S C U8...

Страница 11: ...t update 2022 05 31 IBS Precision Engineering Eindhoven NL 31 40 290 1270 User Manual P11 16 Mechanical detail 17 Power supply Power In 100 240 VAC 50 60 Hz Power Out 12 VDC 1 5 A 18 W CE and UL registered ...

Отзывы: