![IBP KOOL PATCH Скачать руководство пользователя страница 21](http://html.mh-extra.com/html/ibp/kool-patch/kool-patch_instructions-for-use-manual_609321021.webp)
40
41
8. In case problems occur
9. Technical Data
Manufacturer:
Well-Life Healthcare Limited
1Fl., No. 16Lane 454, Jungjeng Rd.
Yunghe City, Taipei City
Taiwan, R.O.C
Type of device:
Device for transcutaneous electrical nerve
stimulation (TENS)
Classification according to 93/42/EWG:
Class IIa
Model:
WL-2301
Maximum permissible ambient temperature:
10°C - 45°C
Maximum permissible relative humidity:
30% - 75%
Maximum permissible air pressure:
700 hPa - 1060 hPa
Battery:
Coin cell Type CR2032, 3 V,
non-rechargeable
Max. output of electrodes:
50 mA
Pulse width:
100 - 150 µs
Pulse rate:
10 Hz - 20 Hz
Operating voltage:
3 V
Weight:
10 g (w/o Battery)
Dimensions (mm):
W:55 x D:10 x H:60
Problem
Possible Cause
Solution
The KOOL
Patch is switched
on, but does not function.
The battery is not inserted.
Insert the battery
.
The battery is empty
.
Replace the battery
.
The KOOL
Patch is defective.
Mail your KOOL
Patch to our
service department, see
Service
and W
arranty
.
or
Dispose of the KOOL
Patch,
see
Disposal and Environmental
Protection
The stimulation is too weak.
The battery is low
.
Replace the battery
.
The patch does not adhere
any longer
.
The patch is dried out.
Replace the patch.