7.1.4 TEMPERATURE COMPENSATION

Listed below are the temperature coefficients used by most major hemodialysis machine and reverse osmosis manufacturers. 
We recommend that you double check this data with documentation from your machine’s manufacturer.

Display

  

To Use For    

 

Temperature Coefficient   

Temperature Coefficient

 

 

 

   at 

25°C 

   at 

37°C

BAXTER  

Baxter (Europe Dialysis Machine)  

2,200 %/K   

 

 

2,200 %/K

BELLCO  

Bellco  

 

 

2,000 %/K   

 

 

2,100 %/K

B.BRAUN  

B.Braun  

 

 

2,000 %/K   

 

 

2,100 %/K

FRESENI  

Fresenius  

 

 

2,000 %/K   

 

 

2,100 %/K

GAMBRO  

Gambro  

 

 

2,000 %/K   

 

 

2,070 %/K

 

 

AK-Type and Phoenix Machines

HOSPAL  

Hospal  

 

 

2,000 %/K   

 

 

2,070 %/K

NIK ALL  

Nikkiso Overall Conductivity  

2,050 %/K   

 

 

2,050 %/K

NIK BIC  

Nikkiso Bicarbonate Conductivity  

2,200 %/K   

 

 

2,200 %/K

STANDARD  

Other Dialysis Machines    

2,000 %/K   

 

 

2,070 %/K

USER TC  

Adjustable  

 

 

0 ... 5,000 %/K 

 

 

0 ... 5,000 %/K

nLF-ISO  

Non Linear Function for 

 

According Standard

 

 

tap, purified and waste water. 

ISO7888 – DIN EN27888

 

 

 

   Reference 

temperature 

25°C

RO-DWA  

Reverse Osmosis DWA 

 

0,00 %/K   

 

 

RO-DWA

 

 

DWA GmbH & Co. KG

 

 

Ubstadter Straße 28

 

 

76698 Ubstadt-Weiher

RO-HERCO  

Reverse Osmosis Herco 

 

2,200 %/K  

 

 

RO-HERCO

 

 

Herco Wassertechnik GmbH  

Reference temperature 25°C

 

 

Planckstraße 26

 

 

71691 Freiberg

RO-LAUER 

Reverse Osmosis Lauer 

 

nLF-ISO according Standard 

 

RO-LAUER   

 

 

LAUER MEMBRAN WASSERTECHNIK 

ISO7888 – DIN EN27888

 

 

GmbH 

 

 

Reference temperature 25°C

 

 

Industriegebiet Speichermatt

 

 

79599 Wittlingen

RO-DWA  

Reverse Osmosis DWA 

 

0,00 %/K   

 

 

RO-DWA

 

 

DWA GmbH & Co. KG

 

 

Ubstadter Straße 28

 

 

76698 Ubstadt-Weiher

RO-HERCO 

Reverse 

Osmosis 

Herco 

 2,200 

%/K 

   RO-HERCO

 

 

Herco Wassertechnik GmbH  

Reference temperature 25°C

 

 

Planckstraße 26

 

 

71691 Freiberg

Содержание Biotech-Lab System

Страница 1: ...www ibpmedical com Biotech Lab User Manual Modular Product Family of Sensors Simulators and Testers HDU Sensor Family pH Temperature Conductivity Pressure Flow...

Страница 2: ...and or this manual on magnetic tape diskette or any other media except for the purpose of own use without written authorization of the IBP Medical GmbH is liable to prosecution Author Werner Pfingstm...

Страница 3: ...Sensor Details 24 8 HDC75 Display and Control Unit 27 8 1 Getting to know the HDC75 27 8 2 Technical Data 28 9 HDU Sensors and Interfaces 29 9 1 HDU CDTP Conductivity Temperature sensor 29 9 2 HDU PRH...

Страница 4: ...lid calibration certificates Only use the Biotech Lab system in a dry environment and do not touch it with wet hands Prevent every mechanical overstraining of electrical wires If the acquired values s...

Страница 5: ...tion and use of the Biotech Lab System The system includes Display and Control Units Sensors Interface Modules Software 1 1 INTENDED USE The Biotech Lab system is intended to be used in biomedical lab...

Страница 6: ...er Products HDC75 Display and Control Unit HDU CDTP Conductivity Temperature Sensor HDU PRH High Accuracy Pressure Sensor Family HDU PRS Pressure Sensor Family HDU FL Flow Sensor Family HDU Pt100 I In...

Страница 7: ...ce PC or Tablet with Microsoft Windows Depending on the requirements you can use the HDC75 or a standard computer with Microsoft Windows as display and control unit for the Biotech Lab modules Biotech...

Страница 8: ...terface for Pt100 temperature sensors Available versions depending on the requirements Temperature Sensor Family Air Surface General Dip HDU YSI400 I Interface for YSI 400 temperature sensors HDU pH I...

Страница 9: ...Biotech Lab server is the central software module Additional App s provide the functionality needed for the different Biotech Lab modules The App s are independent programs which can be used simultane...

Страница 10: ...file in C Users accountname IBP Medical system directory 4 3 INSTALLATION PC For PC s is a special version of the Biotech Lab available Download the software and start setup exe program The different...

Страница 11: ...10 please navigate to http windows microsoft com en us windows 5 1 CHANGING THE DISPLAY LANGUAGE IN WINDOWS 10 To change the default language sign in to Windows 10 and Go to Start Menu and select Sett...

Страница 12: ...You ll see three options Set as default Options Remove Under Options click Download to download the language pack and keyboard for that language If you open Control Panel Languages you ll see the tra...

Страница 13: ...pp allows displaying measuring values from sensors simulators and testers If you start the HDC75 the Lockscreen will look as follows Slide from bottom to top to get the Windows Start Screen Access to...

Страница 14: ...devices and shows the devices suitable for this program During the scan all visibla sensor values are Settings leads to the base settings of BioTechLab Downloads the user manual 6 2 HEAD AREA FUNCTIO...

Страница 15: ...ue display L acc is the accumulated amount in L This is a typical two value display with an additional functionality in the headline Dialysis Machine Standard indicates here Nthe selected conductivity...

Страница 16: ...following functions are available with connected mouse only With a Right Mouse Click into the graphic you get the following menu options More sub functions below leads to Graphic Settings deletes the...

Страница 17: ...e by Mouse To get the following sub functions make a Left Mouse Click followed by a Right Mouse Click Scale Range allows modifying the scales on both sides of the graphic Allows selecting the time int...

Страница 18: ...UES After pressing Minimum Value Maximum Value Arithmetic Mean Value Standard Deviation Current Value Selection of Statistic Interval This is the time range for the calculation of the statistic values...

Страница 19: ...d to 25 C In other words the display is always converted to a solution temperature of 25 C The temperature coefficient which the displayed value is compensated with is expressed as C Unfortunately how...

Страница 20: ...with the machine manual If you have machines from one manufacturer only in your use the temperature coefficient that the dialysis machine is using for compensation If you have different types of dial...

Страница 21: ...e Conductivity 2 200 K 2 200 K STANDARD Other Dialysis Machines 2 000 K 2 070 K USER TC Adjustable 0 5 000 K 0 5 000 K nLF ISO Non Linear Function for According Standard tap purified and waste water I...

Страница 22: ...SENSOR DETAILS The following menu shows the settings for pressure sensors Head area shows the serial number and the date of last calibration Select here the unit you like to see Allows selecting a hi...

Страница 23: ...ng the damping factor 7 4 1 FLOW DIRECTION To avoid malfunction of the sensor please make sure to connect fluid correctly the sensor The sensor has a label showing the flowdirection 7 5 PH SENSOR DETA...

Страница 24: ...nt points Current reading of pH and temperature Allows selecting and inputting the buffer solution value Leads to the next calibration point if any Otherwise the next window shows the calibration resu...

Страница 25: ...by leaking electrolyte can be removed easily by rinsing with reverse osmosis treated water The sample volume should be 100 ml of dialysate or 1 liter of untreated water The sample should be poured in...

Страница 26: ...5 while in proper fast charge mode results in the sensor ports being inaccessible as long as charging is in progress To activate the sensor ports again it is necessary to briefly unplug the USB charge...

Страница 27: ...3 MP fixed focus front facing 2 MP fixed focus rear facing Audio Mono speaker Interface Micro USB B Host also charger connector 5 x RS485 M8 Connector for Biotech Lab Modules Degree of Protection IP54...

Страница 28: ...50 C Conditions Storage Temperature Range 40 to 80 C Operating and Storage Humidity 85 max relative humidity non condensing from 0 to 50 C Dimensions 124 x 34 mm 9 1 2 FLOW THROUGH ADAPTER FOR CONDUC...

Страница 29: ...ly 9 2 1 Technical Data All Versions Reading Readings as mmHg kPa Bar or PSI Measuring Temperature compensated Pressure Connection IBP Sensor Port and G1 8 Female internal thread Membrane HDU PRS vers...

Страница 30: ...dia Compatibility Gasses and Fluids Over pressure 1 5 x full scale HDU PRH50 Range 13 to 50 psi 0 93 to 3 5 bar 700 to 2580 mmHg Resolution 0 1 mmHg Accuracy 0 05 full scale Media Compatibility Gasses...

Страница 31: ...ge Temperature Range 40 to 80 C Operating and Storage Humidity 85 max relative humidity non condensing from 0 to 50 C Dimensions 80 x 34 mm 9 2 2 MEASURING ADAPTER Each HDU Pressure sensor comes with...

Страница 32: ...asured value Repeatability better 0 5 Fluid temperature 20 C to 80 C Pressure 0 1 to 5 bar at 22 C Calibration PTB NIST traceable Communication rate Five times per second Degree of protection IP67 Pow...

Страница 33: ...70 mA via RS485 or USB Interface Environmental Conditions Ambient Operating Temperature Range 0 to 50 C Storage Temperature Range 40 to 80 C Operating and Storage Humidity 85 max relative humidity non...

Страница 34: ...0 to 50 C Dimensions 64 x 45 x 25 mm 9 7 HDU PHPT I Interface for pH Sensor and Pt100 Temperature Sensors pH Range 0 14 pH Resolution 0 01 pH Accuracy 0 02 pH Temperature Pt100 Range 50 100 C Resoluti...

Страница 35: ...US 9 8 HDU YSI400 I Interface for YSI 400 temperature sensors 9 9 HDU YSI700 I Interface for YSI 700 temperature sensors 10 HSI Simulators These modules are under development If you have special needs...

Страница 36: ...it should be not more than one third and preferably one tenth of the permissible error of the confirmed equipment when in use In other words the accuracy of the references to adjust or calibrate the...

Страница 37: ...d solution and calibration verification accuracy Keep solutions tightly capped to avoid evaporation Do not return used solutions to the storage bottle Do not remove solutions from their original bottl...

Страница 38: ...clean any device with any fluids In case of pollution you can wipe the surface of the Biotech Lab devices with a dry and clean cloth 13 3 CALIBRATION PERIOD When delivered new from the factory each de...

Отзывы: