ibidi Gas Incubation System – Silver Line
ibidi Gas Mixer CO
2
/ O
2
Flow
I 9.9
S[ 10.0]
Run
CO2%
Hum%
5.1
5.0
75.2
80.0
Setup
Figure 4: Front view of the Gas Mixer.
The LEDs on the front indicate the status of the channels: connection to the power supply, connection
to the USB, gas control, and humidity control (Figure
Table 4: Control LEDs of the Gas Mixer
LED Mode
Significance
LED off
Channel inactive
LED on
Channel active
LED Air/N
2
/CO
2
blinking
Input gas flow rate very high, please contact
if blinking persists
LED Humidifier blinking
Possible channel error, please contact
On the back of the Gas Mixer, all connections for cables and tubing are provided (Figure
). In the
CO
2
version, the N
2
port is closed inside the instrument and thus non-functional.
The Gas Incubation System takes as input pure CO
2
gas and pressurized air, as well as pure N
2
gas
(only Gas Mixer CO
2
/O
2
). Pressurized air can be supplied by the lab gas line or a small air pump or
compressor. Purity information is given on page
Figure 5: Back view of the Gas Mixer.
Instruction Manual
19