5. To connect the mixed and humidified gas output to the Heated Chamber, screw the Luer Lock
connector into the metal adapter that is fixed in the wall of the Heated Lid.
Figure 12 – Connection of the Heated Tubing Package to the Gas Incubation System.
Caution!
Do not disconnect the Heated Tubing while the Gas Mixer is running. The resulting sudden
change of pressure may lead to damage to the Gas Mixer.
Be careful not to switch the red and blue inputs on the Humdifying Column, as this can also
damage the gas mixer.
5.1.3
Connection of Humidity Sensor Cable
The humidity sensor should be mounted in the Heated Lid of the Heating System (see Section
To operate the Gas Incubation System with humidity control, the humidity sensor cable must be
connected between the Gas Mixer and the humidity sensor.
1. Plug the humidity sensor cable into the Gas Mixer in the connector labeled ”Sensor”. Make
sure that the red dots on the cable and connector line up.
2. Connect the other end to the humidity sensor mounted in the Heated Lid, as shown in Figure
. Again make sure that the red dots on cable and connector line up.