Product: FABian®
Item-No.: 6-6100-260
Instruction Manual, Revision 9 Doc.-No.: 010 Date of Issue: 2020/11/04 Page
17
of
39
1.
To insert the syringe remove it from the packaging and remove existing safety elements. The design
of the syringe and the designation are shown in figure 4.
Figure 4: FABian® Syringe
2.
Insert the syringe into the round opening of the syringe adapter below the syringe lift stage with the
piston facing upwards (see figure 5). Push the syringe as far as possible towards the lid.
3.
Ensure that the syringe piston is correctly placed in the
designated notch. Then push the syringe with the syringe body in the
direction of the housing until it locks into place and until the syringe
body is firmly attached in the adapter (see figure 6).
ATTENTION!
Danger of Damages by Turning the Mini-Column
➔
Do not turn the mini-column to avoid a loosening.
➔
An opening of the syringe can lead to a loss of materials and liqiuds can leak into
the device.
WARNING!
Use Only Original Fabian® Syringes (Without Needle), Risk of Injury
➔
FABian® may be operated exclusively with the original syringe.
➔
It is forbidden to use other syringes equipped with needles. These kinds of
syringes are not applicable for the selection process and can puncture your
fingers and other body parts.
Figure 5: Insertion of the Syringe (1)