Instruction Manual
Страница 1: ...Document No 010 Version 9 Modification Date 2020 11 04 2020 IBA GmbH Instruction Manual...
Страница 2: ...urn Form Decontamination Declaration Revision Index Date Rev 9 2020 11 04 IBA GmbH Bosestr 4 04109 Leipzig Germany Phone 49 0 341 99 38 98 29 Fax 49 0 341 99 38 98 19 E Mail info iba lifesciences com...
Страница 3: ...ssioning Dismantling Disposal 11 4 Transport and Storage 12 4 1 Transport 12 4 2 Storage 12 4 3 Storing and Protecting between Usage Periods 12 5 Prepare FABian for Use 13 5 1 Select Installation Plac...
Страница 4: ...igure 18 Removal of the Screws From the Bottom Part of the Adapter 31 Figure 19 Unscrewing the Screws From the Upper Part of the Adapter 31 Figure 20 Removal of the Syringe Adapter 32 Figure 21 Insert...
Страница 5: ...nual are not being followed the device has been used improperly or has not been used for the specified purpose spare parts accessories or consumables are being used that are not approved by the manufa...
Страница 6: ...disposable products max 11 tubes on a round tube holder and the special FABian syringe are necessary to gain the pure minimal manipulated and non activated target product The operation of FABian is si...
Страница 7: ...to complete by the customer 2 3 Technical Data Table 3 Overview Technical Data Dimensions Weight Mass Capacity Power Measurements Width 198 mm Depth 315 mm Height 230 mm Operation Height 455 mm Weigh...
Страница 8: ...he compliance with maintenance and inspection work as well as the observation of the installation instructions and of the cleaning and maintenance notes Residual Risks Although the device is construct...
Страница 9: ...l be worn 3 5 Safety Instructions for Commissioning The power cable must be routed in such a way that it does not represent a tripping hazard and that no damages can be caused by other devices Use onl...
Страница 10: ...in case of device damages abnormal sounds odors or any other irregularities switch the device off pull out the power plug and call the service technician When unplugging the power plug hold it by the...
Страница 11: ...be subjected to a visual inspection If device damages are visible or the function of FABian is restricted it must be maintained 3 9 Decommissioning Dismantling Disposal FABian should be stopped immedi...
Страница 12: ...uld not exceed 75 Device must not be stored outdoors The floor of the storage room must be flat and dry Do not expose the device to direct sunlight and moisture Storage temperature should be between 1...
Страница 13: ...rations and extreme temperature fluctuations ambient temperature is between 10 C 50 F 35 C 95 F relative humidity of the device environment should not exceed 75 5 2 Prepare Installation To prepare the...
Страница 14: ...p If the ratings for voltage V and maximum current A do not match the device must not be connected 5 Plug the low voltage plug of the power cable in the low voltage socket of the device The low voltag...
Страница 15: ...the power cable in a properly installed and fused power socket 7 Make sure that the power cable is not subjected to tensile or pressure forces DANGER Electric shock Caused by Device Damages or Power...
Страница 16: ...ash green Before first commissioning a check is required to ensure the completeness and functionality and to recognize possible damages This check includes a functional and visual inspection according...
Страница 17: ...he designated notch Then push the syringe with the syringe body in the direction of the housing until it locks into place and until the syringe body is firmly attached in the adapter see figure 6 ATTE...
Страница 18: ...ism 6 2 3 Inserting the Tubes Now you can load the tube holder with the tubes The order of the tubes is already fixed in the manual of the consumable kit and of great importance Please comply with the...
Страница 19: ...ollowing ingredients and tubes are necessary and to arrange as follows 4 x wash solution PBS 2 x biotin 1 x Fab fragment 1 x blood cells Figure 8 Tube Contents If you accidently spill liquids during t...
Страница 20: ...O 1 Start Stop Button Pushing the GO button confirms the program selection the process starts 2 Pause Button Pushing the pause button during operation interrupts the program it continues by pushing th...
Страница 21: ...green see Figure 12 Figure 12 Program A Is Running If you do not start the selected program by pushing the Go button you can choose another program at any time as described above 4 Please bear in mind...
Страница 22: ...that the syringe adapter does not get dirty from the used syringe If this is the case clean the adapter comprehensively see chapter 7 1 3 Remove the loaded tube holder You will find the selected targ...
Страница 23: ...ive fraction is required a mixing up of the reagents from tube 3 8 is necessary for an optimum use After the syringe and the tubes are removed you can switch off the device or insert new consumables f...
Страница 24: ...unusual situations or errors occur please contact the IBA GmbH Do not open the device or carry out any repair and maintenance without the permission of the IBA GmbH When opening the device housing an...
Страница 25: ...ean it with a thin cloth or rather a thin brush which will be pass through the adapter 6 It must be ensured that no liquids leak into the inside of the housing cover Under no circumstances power socke...
Страница 26: ...packaging Make sure that FABian has been cleaned and disinfected extensively The device has to de dry and clean We recommend a conclusion of a maintenance agreement in which the maintenance intervals...
Страница 27: ...10 ml standard IBA GmbH 1 Disposable The consumables have to the disposed of according to the guidelines for autoclavable laboratory waste 1 Tubes Figure 14 Image of a Tube The tubes are disposables...
Страница 28: ...00 000 IBA GmbH Service continuous 2 PSU 6 6101 000 IBA GmbH Service continuous 3 Power Cable 6 6102 000 IBA GmbH Service continuous 4 Preset SD Card 6 6103 000 IBA GmbH Service continuous 5 Instructi...
Страница 29: ...ery of FABian It serves as a holder for a maximum of 11 tubes and is to be locked to the roundabout The notch on the roundabout guarantees the accurate locking of the tube holder If the tube holder is...
Страница 30: ...yringe volume a corresponding syringe adapter is necessary Please refer to table 7 for possible adapter variations The replacement of the adapter is demonstrated in chapter 9 Please use exclusively or...
Страница 31: ...se refer to the following steps for the exchange of the syringe adapter Exchange of the Syringe Adapter For the exchange of the syringe adapter it is necessary to dismantle the old adapter 1 Remove th...
Страница 32: ...s to install the bottom part see figure 22 6 Insert the upper part of the adapter into the notch of the device head with the bar facing upwards Close the cover Take care that the bar of the upper part...
Страница 33: ...sion 9 Doc No 010 Date of Issue 2020 11 04 Page 33 of 39 7 Now tighten the screw below the upper part of the adapter by using the allen wrench see fig 24 Figure 23 Insertion of the Upper Part of the A...
Страница 34: ...ideline 2012 19 EU Waste of Electrical and Electronic Equipment WEEE Pursuant to this regulation it is not allowed to dispose of the device with normal household garbage This is indicated by the follo...
Страница 35: ...are compatible with the PSU 2 The power cable is undamaged and fully functional 3 The PSU is undamaged and fully functional 4 All protection devices housing plugs etc are proper and free of defects 6...
Страница 36: ...Product FABian Item No 6 6100 260 Instruction Manual Revision 9 Doc No 010 Date of Issue 2020 11 04 Page 36 of 39...
Страница 37: ...Product FABian Item No 6 6100 260 Instruction Manual Revision 9 Doc No 010 Date of Issue 2020 11 04 Page 37 of 39...
Страница 38: ...Product FABian Item No 6 6100 260 Instruction Manual Revision 9 Doc No 010 Date of Issue 2020 11 04 Page 38 of 39...
Страница 39: ...Product FABian Item No 6 6100 260 Instruction Manual Revision 9 Doc No 010 Date of Issue 2020 11 04 Page 39 of 39...