1. User Dead Update 

1.1. Method 

1) Launch the application, Thermal Expert. 

2) Put on the lens cap. 

3) Push the ‘Image calibration’ button for 8 seconds. 

4) The pop-up message, ‘Dead Update’, will be shown. 

5) Touch the ‘Start’ button. 

6) Repeat the routine, 3) ~ 5), until the dead pixels are not appeared. 

    (2 or 3 times is recommended) 

 

     

1.2. Details 

1) The updated dead pixel is saved as a file in a smart phone. 

2) The path of a saved file is i3cam/dead.txt 

3) Even if the thermal expert camera is re-plugged, the updated dead file is   

applied. 

4) If the dead file is deleted on your smart phone, then you need to re-update   

dead pixel. 

 