IACER Srl
20
MNPG117-02
If the body surface is very irregular, making it difficult to obtain good contact
between the ultrasound head and the body (e.g. ankle or foot), or if direct
contact must be avoided (e.g. due to pain), the affected area may be treated
under water (subaqual method). Use two liters of water, with a temperature
not higher than 25 ° C. The water should be degassed (by previous boiling) in
order to prevent air bubbles that could decrease the effectiveness of the
treatment.
WARNING
. Never apply the gel to the ultrasound head. The treatment
head will register this as contact and may emit ultrasound energy,
which could damage the ultrasound head. Always use the gel
certificated with the requirements of the medical, such as with CE
mark.
WARNING. The device can be used to perform immersion
treatments. The handle and its cable are protected from damage by
water infiltration according with the gradeIPX7.
Programs features and main applications
Refer to the following table for programs features. All parameters are
adjustable by the user.
PROG.
Medical
Progr.
Yes/No
FREQ.
DUTY
CYCLE
TIME
SUGGESTED
INTENSITY
U-01
Yes
1MHz
80%
10min
1.0W/cm²
U-02
Yes
1MHz
50%
10min
1.0W/cm²
U-03
Yes
1MHz
50%
20min
1.5W/cm²
U-04
Yes
1MHz
50%
15min
1.0W/cm²
1.5W/cm²
2.0W/cm²
U-05
Yes
3MHz
80%
15min
1.0W/cm²
U-06
Yes
1MHz
30%
15min
1.5W/cm²
U-07
Yes
1MHz
80%
15min
1.0W/cm²
1.5W/cm²
U-08
Yes
1MHz
80%
8min
1.5W/cm²
U-09
Yes
1MHz
50%
12min
1.5W/cm²
U-10
Yes
3MHz
80%
10min
1.0W/cm²