IACER Srl
24
MNPG69-09
Spare parts
The manufacturer shall make available the original spare parts for the
equipment at any time. To request them:
I.A.C.E.R. S.r.l.
Via Enzo Ferrari, 2
• 3003
7 Scorzè (VE)
Tel. 041.5401356 • Fax 041.5402684
For the purpose of maintaining the warranty, the functionality and safety of
the product it is recommended to use only original spare parts supplied by
the manufacturer (also consult the
Warnings
paragraph).
Interference and elctromagnetic compatibility tables
The MAG700 equipment has been designed and manufactured according to
the TECHNICAL STANDARD on ELECTROMAGNETIC COMPATIBILITY
legislation EN 60601-1-2:2015 with the aim of providing adequate
protection from harmful interference when installed in homes and health
establishments.
It Is advisable to use the device at a distance of at least 3 meters from
televisions, monitors, mobile phones, WI-FI devices or any other electronic
equipment as such equipment may affect the operation of the device.
In particular, portable communication equipment such as WI-FI devices, mobile
phones, cordless phones and their base stations, walkie-talkie, can affect the
medica
l device and it’s recommended a separation distance
𝑑
calculated from
the fabricant in table
RF IMMUNITY ASPECTS
, column
800𝑀𝐻𝑧 − 2,5𝐺𝐻𝑧
,
paragraph EMC tables. Example: for a mobile phone with
2𝑊
maximum output
power the separation distance
𝑑 = 3,3𝑚
in order to obtain an immunity level
of
3𝑉/𝑚
or a separation distance
𝑑 = 0,5𝑚
for an immunity level of
20𝑉/𝑚
..
The device must be installed and commissioned in compliance with the
information on electromagnetic compatibility supplied in this manual. Also
consult the
EMC Tables
paragraph.
Using accessories, transducers and cables other than those specified, except
for those transducers and cables sold by the manufacturer as spare parts for
internal components, may result in increased emissions or decreased
immunity of the device.
The device should not be placed next to or on top of other devices. Should
it prove necessary to place it next to or on top of other devices, supervision
is essential at all times to control its normal functioning.
Содержание MAG700
Страница 1: ...USER MANUAL Magnetotherapy model MAG700...
Страница 2: ......
Страница 31: ......
Страница 32: ......