MNPG127-02
Pag. 6
! WARNINGS!
-
The perfect functionality of the device is guaranteed in accordance with the
rules of installation and of use included, only with original accessories and
spare parts.
-
If there are problems or installation difficulties, please contact the I.A.C.E.R.
srl technical assistance department.
-
The correct position while moving the machine: the apparatus has to be
moved exclusively by gripping it with both hands on the curved profiles of
the lid.
-
If you want to install an extern interlock circuit, contact exclusively qualified
technicians and supply them the scheme correspondent to the room used for
the emission of the treatment. A bad installation of the device can to
generate serious ocular lesions.
-
Before connecting the cable to the mains plug, check that the equipment
wasn’t damaged during transport. Ensure that the power supply
specifications on the mains socket correspond with the information on the
label attached to the back of the unit.
-
The electric current that powers the unit is VERY DANGEROUS. Before
connecting or disconnecting the power cable from the connecter on the unit,
make sure it is plugged out from the mains socket.
-
The power cable has an earthed plug for safety reasons.
-
Only use with a mains socket suitable for use with earthed systems.
-
The equipment should only be connected to electrical systems that fully
comply with
regulations.
-
Check the integrity and the existence of ground conductor if extensions
cables are used.
-
Connect the equipment directly to the wall socket without using extensions.
Failure to comply with these warnings may result in dangerous electrical
discharges that could cause injury operators and compromise the functioning
of the unit.
USE
-
The laser therapy treatments must be provided, under the strict control of
the operator, patients conscious, able to interact with the operator in
response to stresses transmitted by the device; in case of default to the
indications given, I.A.C.E.R. srl shall not be consider responsible for any
accidents .
-
The use of the controls or regulations or the execution of different
procedures from those specified in this user manual can cause the exposition
to dangerous radiation.
-
The operator has the responsibility to verify
that the issuing head remains
well in contact to the zone of treatment, to avoid the emission to different
zones from those to be treated.
-
The laser radiation that outgoings from the device is dangerous: always use
the appropriate glasses. Always avoid the exposition of the eyes to the direct
or reflected laser beam.
-
It is recommended not to start treatment if the machine is not in perfect
mechanical condition or if the laser does not present characteristics of
approved for this purpose (see the specifications table).
-
During the delivery the handpiece must be positioned in contact with the
part to be treated. After activating the handpiece through the pedal contact,
avoid that it moves or is directed to different areas.
-
THE PROBE MUST NEVER BE DIRECTED TO AREAS OF BODY SENSITIVE TO THE
LASER RADIATION, FOR EXAMPLE THE EYES.
-
ALWAYS AVOID THE EXPOSITION OF THE EYE TO THE DIRECT OR REFLECTED
LASER BEAM.
-
Do not leave the device switched on unattended, always switch off after use.
-
Use only probes accurately cleaned and disinfected after each treatment in
order to avoid environment and users contamination.
The operator must pay attention to the necessity of a periodic maintenance
(every 2 years) of the probes/applicators. IACER srl authorised personnel
should carry out such operations.
It is absolutely forbidden in presence of anaesthetic inflammable substances
and in full oxygen environment. I.A.C.E.R. will be not responsible if these
indications will be not complied.
-
It is absolutely forbidden to cover the ventilation slots: such an action may
not allow the machine to work in safe conditions. In case of non-compliance
with this indication, I.A.C.E.R.Srl will not be responsible for any accidents.
-
It’s important to pay the attention of the operator to the necessity to verify
the correctness of the electric installation of device before activating the
supply switch.
-
It is advisable to suspend the therapeutic treatment if it were to appear some
disturbances during its emission.
-
It’s strongly advised not to hold the device on in state of start without using
the probe, it could overheat.