ELECTROMAGNETIC INTERFERENCES
The I-TECH LA500 equipment has been designed and manufactured according to the
ELECTROMAGNETIC COMPATIBILITY DIRECTIVE 2004/108/CE with the aim of
providing adequate protection from harmful interference when installed in homes
and health establishments.
The I-TECH.AR equipment does not generate significant radio frequency energy and is
adequately immune to radiated electromagnetic fields. Therefore it does not
detrimentally interfere with radio-electric communications, electro-medical
equipment for monitoring, diagnosis, therapy and surgery, office electronic devices
such as computers, printers, photocopiers, fax machines, etc. or
any electric or electronic equipment used in these environments, as long as said
equipment complies with the ELECTROMAGNETIC COMPATIBILITY directive.
In any case, in order to avoid any interference problems, we recommend that you
operate the therapy equipment far enough away from critical equipment for
monitoring vital patient functions, and that you be careful when applying therapy to
patients with pacemakers.
TROUBLESHOOTING CHART
PROBLEM
POSSIBLE CAUSE
SOLUTION
The LCD display in
the frontal panel
does not turn on:
the device does
not work.
The power plug is inserted not
correctly in the socket.
Check that the socket works properly.
The power cable is not inserted
correctly into the connector of the
device.
Insert correctly the plug and the cable
in the connector of the device.
The power cable is worn out or
interrupted
Replace the power cable.
The emergency switch is off.
Turn on the emegency switch.
Fuse/s is/are defective or
interrupted.
Replace missing, defective or
interrupted fuse/s
The control electric circuit is
broken.
Contact IACER srl service centre.
The LCD display in
the frontal panel
does not turn on.
Defective components in the
control electric board.
Contact IACER srl service centre.
Some controls on
the front control
panel do not work
properly.
Faulty keys or buttons.
Contact IACER srl service centre.
Control electronic circuit failure.
The device does
not emit the
therapy.
Parameters
are
not
setted
correctly.
Check that the parameters have been
setted correctly.
Laser sources not working or
depleted.
Check laser source emission mode.
Faulty components in the control
electronic circuit.
Contact IACER srl assistance centre.
Faulty supply of laser sources.
The unit works
properly but with
a significant
decrease in
efficiency of the
treatment.
Faulty or depleted laser source.
Contact IACER srl assistance centre.
Possible break down in power
generator circuit of the unit.
The equipment
starts up, or
seems to work
properly, but
there is no
emission.
No safety key or the interlock
circuit is open.
Insert the DIN safety key into the
front socket. Reset the safety
conditions.