Pulse Oximeter  

FOX-200 (

JPD-500E(LED)) 

Issuing Date: April, 2020 

   

     

 MNPG302-00 

   Instructions to User 

 

 

Dear users, thank you very much for purchasing the Pulse Oximeter. 
This Manual is written and compiled in accordance with the council directive 
MDD 93/42/EEC for medical devices and harmonized standards. In case of 
modifications and software upgrades, the information contained in this 
document is subject to change without notice. 
Please read the User Manual carefully before using this product. 

 

WARNING: 

■  Uncomfortable or painful feeling may appear if using the device 
ceaselessly, especially for the microcirculation barrier patient. It is 
recommended that the sensor should not be applied to the same finger for 
over 2 hours. 

■  The light (the infrared is invisible) emitted from the device is harmful to the 
eyes, so the user and the maintenance man should not stare at the light. 

■  Enamel or Acrylic fingernail polish or other fingernail applications may 
distort and/or produce inaccurate readings. 

■  Please refer to the correlative literature about the clinical restrictions and 
caution. 

 

   1 Safety 

 

 

1.1 Instructions for Safe Operations 

■  Check the main unit and all accessories periodically to make sure that 
there is no visible 

damage that may affect patient’s safety and monitoring 

performance about cable and transducers. It is recommended that the device 
should be inspected once a week at least. When there is obvious damage, 
stop using the monitor. 

■  Necessary maintenance must be performed by qualified service 
engineers ONLY. Users are not permitted to maintain it by themselves. 

■  The oximeter cannot be used together with devices not specified in User 
Manual. Only the accessory that appointed or recommendatory by 
manufacture can be used with this device. 

■  This product is calibrated before leaving factory. 

1.2 Warning 

■  Explosive hazard---DO NOT use the oximeter in environment with 
inflammable gas such as some ignitable anesthetic agents. 

■  DO NOT use the oximeter while the user is measured by MRI and CT. 

■  The person who is allergic to rubber cannot use this device. 

■  The disposal of scrap instrument and its accessories and packing 
(including battery, plastic bags, foams and paper boxes) should follow the 
local laws and regulations. 

■  Please check the packing before use to make sure the device and 
accessories are totally in accordance with the packing list, or else the device 
may have the possibility of working abnormally. 

1.3 Attentions 

●  Keep the oximeter away from dust, vibration, corrosive substances, 
explosive materials, high temperature and moisture. 

●  If the oximeter gets wet, please stop operating it. 

●  When it is carried from cold environment to warm or humid environment, 
please do not use it immediately. 

●  High temperature or high-pressure steam disinfection of the oximeter is 
not permitted. Refer to User Manual in the relative chapter for instructions of 
cleaning and disinfection. 

●  Do not have the oximeter immerged in liquid. When it needs cleaning, 
please wipe its surface with medical alcohol by soft material. Do not spray 
any liquid on the device directly. 

●  Do not use the device on infant or neonatal patients. 

●  The product is suitable for children above four years old and 
adults (Weight should be between 15kg to 110kg). 

Содержание FOX-200

Страница 1: ...rs ONLY Users are not permitted to maintain it by themselves The oximeter cannot be used together with devices not specified in User Manual Only the accessory that appointed or recommendatory by manuf...

Страница 2: ...2 Feature Operation of the product is simple and convenient The product is small in volume light in weight total weight is about 50g including batteries and convenient in carrying Power consumption of...

Страница 3: ...fluorescent lamp dual ruby light infrared heater direct sunlight etc 4 Strenuous action of the subject or extreme electrosurgical interference may also affect the accuracy 4 Accessories 1 One hanging...

Страница 4: ...air or clean it by dry clean fabric Using the medical alcohol to disinfect the product after use prevent from cross infection for next time use Please take out the batteries if the oximeter is not us...

Страница 5: ...rmation Display Mode Display Format LED display The Pulse Oxygen Saturation SpO2 Digital Pulse Rate PR Digital Pulse Intensity bar graph Digital bar graph display SpO2 Parameter Specification Measurin...

Страница 6: ...ply network that supplies buildings used for domestic purposes Warranty Your product is warranted to be free of defects in materials and workmanship for one year from the original purchase date The de...

Страница 7: ...nen direkt von der Bildschirmanzeige 5 Dr cken Sie im Boot Strap Zustand die Taste und das Ger t wird zur ckgesetzt Fingern gel und Lumineszenzrohr sollten auf derselben Seite liegen Fehlersymptome Ur...

Страница 8: ...s l ger utilisation portable et op ration simple cran LED visible Indication de basse tension de la batterie L appareil s teindra automatiquement 16 secondes plus tard si aucun signal n est d tect L a...

Страница 9: ...illez contacter le centre de service local L cran s teint soudainement 1 L appareil s teint automatiquement en l absence de signal dans les 16 secondes 2 Les batteries sont presque puis es 1 Normal 2...

Страница 10: ...funzioni correttamente LaSpO2 e il battito cardiaco non vengono visualizzati stabilmente 1 Il dito non posizionato abbastanza in fondo 2 Il dito o il paziente si sta muovendo 1 Posizionare il dito co...

Страница 11: ...Posible raz n Soluci n La SpO2 y la frecuencia del pulso no se muestra normalmente 1 El dedo no est colocado correctamente 2 El valor de la SpO2 en el paciente es demasiado bajo para ser detectado 1 C...

Страница 12: ...71 Xintian Road Fuyong Street Baoan Shenzhen Guangdong China E mail info jumper medical com Tel 86 755 26692192 26696279 Web www jumper medical com Distribuited by I A C E R Srl Via S Pertini 24 a 30...

Отзывы: