6.4.3
Draining the machine
1.
Put the i-cover® 2.5 on a table with the
backside towards you.
2.
Release the drainage tube (Figure 22, pos 1)
via the clips (Figure 22, pos 3).
3.
Put the end of the drainage tube into a
proper container for disinfectants.
4.
Remove the yellow stopper from the fluid
tank (Figure 12, pos 2).
5.
Open the tap (Figure 22, pos 2).
6.
When empty close the tap (Figure 22, pos 2).
7.
Replace the yellow stopper
(Figure 12, pos 2).
8.
Replace the drainage tube (Figure 22, pos 1)
using clips (Figure 22, pos 3).
Figure 22