38
www.i-sens.com
www.i-sens.com
39
Step 4
On pressing the
▲
button , the number
indicating the minute will start blinking.
Press the
▼
button to set the minute.
Step 5
Press the
S
button to finish and to go to
'alarm 2' mode.
Repeat steps 2 to 5 to set the remaining
time alarms (alarm 2–5).
Step 6
Press the
S
button for 3 seconds to finish
and turn the meter off.
Message
What It Means
What To Do
Understanding Error Messages
Maintenance
A used test strip
was inserted.
The blood or control
solution sample was
applied before the
symbol appeared.
The temperature
during the test was
out of the
operating range.
Repeat the test with a
new test strip.
Repeat the test with a
new test strip and
wait until the
symbol appears
before applying the
blood or control
solution sample.
Move to an area where
the temperature is
within the operating
range (41–113 ˚F/
5–45 ˚C) and repeat the
test after the meter and
test strips have reached
a temperature within
the operating range.
Содержание VetMate