Step 9
Obtain a blood sample using the 

lancing device. Place the device against 

the pad of the finger as shown in the 

diagram. Press the release button. 

Remove the device from the finger. Wait 

a few seconds for a blood drop to form. 

A minimum volume of 0.5 microliter is 

needed to fill the confirmation window. 

(Actual size of 0.5 µL :    )
Step 10
After the  

 symbol appears on the screen, apply the blood 

sample to the narrow end of the test strip till the meter beeps. 

If the confirmation window is not filled in time because of 

abnormal viscosity (thickness and stickiness) or insufficient 

volume, the Er4 message may appear.

It is recommended that the application of blood sample to the  

test strip be performed virtually vertical to the sample site as 

shown in the diagram below.

Applying Blood Sample

3

4

5

6

7

Good Sample

Insufficient 

Sample

Note:

 The meter may switch off if the blood sample is not 

applied within 2 minutes of the 

symbol appearing 

on the screen. If the meter turns off, remove the strip and 

reinsert it and apply blood sample after 

symbol 

appears on the screen.

Step 11
After the blood sample has been 

applied, the display segments will 

rotate clockwise on the meter display.
The test result will appear after the 

meter counts down from 5 to 1. The 

result will be automatically stored in 

the meter’s memory.
If the test strip is removed after the test 

result has been displayed, the meter 

will automatically switch off after 3 

seconds. Discard used test strips safely 

in appropriate containers.