S570/3
CALIBRATION
To calibrate this meter, you will require a calibrated measure greater than
5L in volume. A 20L measure can is recommended and is used in the
below example:
1. Wait for the meter to go into standby mode.
2. Press the
RESET
button so that the resettable totaliser resets to zero.
3. Start dispensing fluid into the calibrated measure can at full flow until
desired calibrated level is reached, i.e.20.00L.
4. Now press and hold the
MENU
button until
“Cal”
is displayed in the
bottom left of the screen. This is the calibration input screen. The first
digit will now flash.
5. We now need to input
020.00L
into the display. Press the
RESET
button to select the correct digit. Once the correct digit is displayed
press the
MENU
button to scroll to the next digit. Continue until
020.00L
is displayed.
6. Now press and hold the
MENU
button to exit the calibration menu.
7. Carry out a test measure to ensure that the new calibration setting is
accurate.
Please see manufacturer’s instructions for additional technical information,
exploded diagrams, troubleshooting guide etc.
REPLACING THE BATTERY
You will need to remove the 4 screws on the rear of the meter to remove
the display and gain access to the battery compartment. The replacement
CR2 battery part code is
FM.DA1.BATT