iRIS 350FX Reference Manual V1.50
50
HyQuest Solutions (NZ) Ltd
- PO Box 9466, Hamilton, New Zealand Tel: +64 7 857-0810 Email: [email protected]
7.2.5.4
System Status 3 Screen (Level 1)
The System Status 3 screen is always available, no matter what level of access has been selected. The
information shown on this screen includes the internal battery voltage, supply voltage, internal temperature
and if applicable the RSSI (signal strength).
7.2.5.5
System Status 4 Screen (Level 1)
The System Status 3 screen is always available, no matter what level of access has been selected. The
information shown on this screen includes the voltage being measured at the AI1 – AI4 terminals in volts.
These values are useful for checking input signals and also for the calibration process.
7.2.5.6
System Status 5 Screen (Level 1)
The System Status 5 screen is also always available, no matter what level of access has been selected. The
information shown on this screen includes the current status of the digital channels whether they are an
input (DIx:y) or an output (DOx:y) (where x is the channel and y is the state 0=OFF, 1=ON). This screen also
shows the current alarms status (A) which is 1 if any alarms are active and also the currently selected
communications scheduler interval in minutes. Finally it displays the Start Of Data (SOD) and End Of Data
(EOD) pointer values.
7.2.5.7
Logging Control Screen (only accessible when logged in)
The Logging Control screen is a special screen that is only available on level 1 when a user is logged in. It can
be used to temporarily disable logging when the logged in user is making changes to or testing sensors and
does not want to have the data logged. It is reached by pressing the + key from Status Screen 4. Use the Enter
key to toggle logging on or off. If the logging is disabled, it is always re-enabled automatically when the user
logs out, either manually or on an inactivity timeout (30 minutes).
Use the – or + keys to switch the logging on or off. Pressing the <Enter> key will return the display to the
top-level Welcome (System Status 1) screen.
Battery: 12.57 V
Supply: 23.78 V
Temp: 21.6
C
RSSI: -75 dBm
-
Move back to System Status 2 screen
+
Move forward to System Status 4 screen otherwise unused.
Enter
Move down to Main Menu screen
Alt Enter
Log in (If PIN code <> 0), otherwise unused
DI1:0 DI2:1 A: 0
DI3:0 DO4:0 I: 360
SOD Ptr: 177104
EOD Ptr: 177336
-
Move back to System Status 3 screen
+
Move to Logging Control screen if logged in. Otherwise unused.
Enter
Move down to Main Menu screen
Alt Enter
Log in (If PIN code <> 0), otherwise unused
LOGGING CONTROL
Use Enter key to
Enable/disable.
=> Enabled
-
Move back to System Status 4 screen
+
Unused
Enter
Toggle logging on or off
Alt Enter
Unused
AIN1: 1.7541 V
AIN2: 0.6821 V
AIN3: 1.2390 V
AIN4: 0.0006 V
-
Move back to System Status 3 screen
+
Move forward to System Status 5 screen
Enter
Move down to Main Menu screen
Alt Enter
Log in (If PIN code <> 0), otherwise unused