background image

 

HyQual Multi-probes User Manual v3, 08.07.2022 
KISTERS Controlled, Template: Word-with-table-of-contents-Template-Company.docx, v1.1 

Page 10 / 23 

 

5.3

 

Set Time and Date 

To calibrate Time and Date, open the Calibration menu, select “Set Time and Date” to see 

the HyQual 

probe

’s 

current time and date.  If you wish to change any of those values, just type the new value in the 

appropriate box or click the box at the bottom of the screen.  If you wish to synchronize the HyQual 
probe 

time and date with that of your Display, click the “synchronize” box.

 

 

Operation

 

6.1

 

Sensor Warm Up 

The HyQual probe knows the warm-up times required for all the sensors you have enabled.  It figures 
out exactly when to turn the various sensors on so that a frame of data can be taken exactly at the cor-
rect time.  For instance, the DO sensor takes 20 seconds to warm up and the turbidity sensor takes 25 
seconds to warm up.  So, if you have only a DO sensor, the warm-up time will be 20 seconds.  If you 
have DO and turbidity sensors, the warm-up time will be 25 seconds. 

6.2

 

Four Basic Deployment Methods 

There are four basic deployment methods for The HyQual probe. 

6.2.1

 

Manual Data Collection 

Manual Collection is also known as profiling, surveying, spot measurement, site-to-site measurements, 
etc., means you are present at the monitoring site and using a Display for observing measurements.  
This allows you to make data-based decisions in the field in real-time at multiple monitoring sites in one 
day.  The Display can be a laptop, or almost any tablet or smart phone.  You can record measurements 
using the HyQual probe

’s Snapshot or Automatic Snapshot features.  You might “snapshot” a series of 

measurements in one or more lakes or streams during the day, and then download your data to your 
desktop PC that evening.  If your Display has email, you can email the data to anyone you wish, including 
yourself. 

6.2.1.1

 

Uploading Snapshot and AutoSnap Data 

If you are using a phone or tablet as your display, Snapshot and AutoSnap data are stored in your Dis-
play.  If you wish to upload that data to a PC, simply follow the same procedure you would use to syn-
chronize your PC and phone or tablet to transfer pictures, contact lists, etc. 
You can also email data files from your phone or tablet if they have Web access. 

6.2.2

 

Unattended Logging 

Unattended Logging means that you have set the HyQual probe into its Logging mode, deployed the 
HyQual probe in the proper location in the water, and left the site.  The HyQual probe can run for weeks 
at a time with cable-supplied power or an optional HyQual Internal Lithium Battery Pack.  You can, for 
instance, set the instrument to take a set of readings every half-hour, anchor it in an estuary, and return 
after two weeks to retrieve the instrument and download the data to a PC, laptop, tablet, or phone. 

6.2.2.1

 

Log Start on Power Up 

HyQual starts logging when the user activates logging (see 5.1), and then provides power.  
When power is first provided, the red LED will blink five times to confirm that Logging is activated, and 
the green LED will blink briefly to confirm that the HyQual probe is receiving adequate voltage to start 
Logging.  

Содержание HyQual

Страница 1: ...HyQual Water Quality Multi Probe HyQuest Solutions KISTERS Group 2022 05 03...

Страница 2: ...ction 10 6 2 2 Unattended Logging 10 6 2 3 Telemetry Relay 11 6 2 4 On Line Monitoring 12 7 1 When to proceed to maintenance 12 7 2 Maintenance 1 Keep your device clean 12 7 3 Maintenance 1 Calibratio...

Страница 3: ...ilidad de tales da os Cualquier error encontrado en cualquier producto de HyQuest Solutions debe ser comuni cado a HyQuest Solutions donde se har todo lo posible para resolver r pidamente el pro blema...

Страница 4: ...industrial specification that defines the electrical interface and physical layer for point to point communication of electrical devices The RS 485 stand ard allows for long cabling distances in elec...

Страница 5: ...s battery operates only at the surface and its job is to power the multiprobe while you re pairing the instrument with a data display via Bluetooth wireless technology A USB port on a PC is another wa...

Страница 6: ...send the probe back to us at your own expense Additional sensors Barometric pressure is calculated with the use of the depth sensor therefore you will need to add an optional depth sensor to any of yo...

Страница 7: ...ufacturer s named repair center You can find information on how to re turn items for repair in the relevant section of the HyQuest Solutions website Disposal instructions After taking any HyQuest Solu...

Страница 8: ...ufficient water flow that can be easily ac cessed for maintenance and protected from external elements such as animals and items such as stones or sticks that may damage the probe 4 1 Protection for u...

Страница 9: ...n to toggle between Logging ON and Logging OFF The Setup function for Logging lets you name your Logging file and set the logging interval time be tween logged lines of data Figure 1 Switch Logging ON...

Страница 10: ...The Display can be a laptop or almost any tablet or smart phone You can record measurements using the HyQual probe s Snapshot or Automatic Snapshot features You might snapshot a series of measurements...

Страница 11: ...rchase this battery after as it is not integrated to the probe Please note that this battery ads several inches to the length of the probe and that it can also be used with the HyQual 200 only that it...

Страница 12: ...ed in the control loop or decision making process Maintenance 7 1 When to proceed to maintenance Judgment gained from observing your field conditions and data requirements provides information about w...

Страница 13: ...tuation for which the correct parameter value is known Here s the general procedure 1 Clean the sensor and perform any necessary sensor specific maintenance 2 Select a calibration standard whose value...

Страница 14: ...c But you can elect to accept any SRF 7 3 3 Choosing Calibration Standards For best results choose a calibration standard whose value is close to what you expect to see in the field For example calibr...

Страница 15: ...ate of the calibration the parameter calibrated the reading before the calibration was accepted the read ing after the calibration was accepted the raw sensor reading the SRF and a few other details I...

Страница 16: ...follow the calibration instructions on the screen The manufacturers of optical DO sensors usually recommend that you not calibrate the zero DO point because the zero DO point drift is very low However...

Страница 17: ...don t report because it has little meaning in natural waters You can t calibrate TDS or salinity directly because they are calculated from Conductivity You can however calibrate TDS with a TDS standar...

Страница 18: ...ous circle at the end of the sleeve This not only purges bubbles from the electrolyte but also cleans nasty stuff out of the junction Figure 4 Filling in fresh pH reference electrolyte 7 3 10 ORP ORP...

Страница 19: ...s a small hole in the back of the transducer There is a tube inside the cable i e a Vented Cable to connect the Stage sensor to the atmosphere so that changes in barometric pressure will not affect t...

Страница 20: ...e small set screw on the wiper arm 2 Remove the wiper pad from the wiper arm and replace the pad 3 Place a new wiper arm on the motor shaft so that the set screw faces the flat spot on the motor shaft...

Страница 21: ...bor charge 1 2 days down time Repair HyQuest Solutions precision instruments and data loggers are produced in quality controlled processes All HyQuest Solutions production and assembly sites in Austra...

Страница 22: ...lay devices not for eld use calibration solutions copper anti fouling kit USB adapter connect DB9 to USB inte grated SDI 12 and MODBUS output SDI 12 adapter cable MODBUS adapter cable and Android disp...

Страница 23: ...ion battery must be dis posed of according to legal provisions EU WEEE Directive As players in the environmental market KISTERS AG and HyQuest Solutions are committed to supporting efforts to avoid a...

Отзывы: