General Notes
Medical electrical equipment requires special precautions regarding
electromagnetic compatibility (EMC) and needs to be installed and put into
service according to the EMC information provided in this document and
the remainder of the instructions for use of this electronics box. Portable
and mobile RF communications equipment can affect medical electrical
equipment. Cables and accessories not specified within the instructions
for use are not authorized. Using other cables and/or accessories may
adversely impact safety, performance, and electromagnetic compatibility
(increased emission and decreased immunity). Care should be taken if the
equipment is used adjacent to or stacked with other equipment; if adjacent
or stacked use is inevitable, the equipment should be observed to verify
normal operation in the configuration in which it will be used.
Electromagnetic Emissions
This equipment is intended for use in the electromagnetic environment
specified below. The user of this equipment should assure that it is used in
such an environment.
Electromagnetic Immunity
This equipment is intended for use in the electromagnetic environment
specified below. The user of this equipment should assure that it is used
in such an environment.
Immunity
Against
IEC 60601-1-2
Test Level
Compliance
Level (of this
electronics
box)
Electromagnetic
Environment
Electrostatic
discharge, ESD
(IEC 61000-4-2)
Contact Discharge:
± 8 kV
Air Discharge:
± 15 kV
± 8 kV
± 15 kV
Floors should be wood,
concrete, or ceramic
tile. If floors are covered
with synthetic material,
the relative humidity
should be kept at levels
to reduce electrostatic
charge to suitable
levels.
Electrical fast
transients/
bursts
(IEC 61000-4-4)
Power Supply
Lines: ± 2 kV
Input DC Power
Ports ± 2 kV
Signal Input/Output
Lines: ± 1 kV
± 2 kV
± 2 kV
± 1 kV
Mains power quality
should be that of a
typical commercial or
hospital environment.
RF Proximity
(IEC 61000-4-3)
385 Pulse
Modulation, 18 Hz
450 FM + 5Hz
deviation: 1 kHz
sine
710
745
780 Pulse
Modulation: 217 Hz
810
870
930 Pulse
Modulation: 18 Hz
1720
1845
1970 Pulse
Modulation: 217 Hz
2450 Pulse
Modulation: 217 Hz
5240
5500
5785 Pulse
Modulation: 217 Hz
27 V/m
28 V/m
9 V/m
28 V/m
28 V/m
28 V/m
9 V/m
Equipment with high
RF emissions should
be kept at a distance to
reduce the likelihood of
interference.
Surges on AC
mains lines
(IEC 61000-4-5)
Common Mode:
± 2 kV
Differential Mode:
± 1 kV
± 2 kV
± 1 kV
Mains power quality
should be that of a
typical commercial or
hospital environment.