7.3

 

Luminometer Error Codes 

During  normal  operation,  the  luminometer performs 

various  self-checks  on  its  internal  components.  If  a 

problem  is  detected,  the  display  will  show  an  error 

number: 

Error Code

 

Possible Causes

 

E1

 

Not applicable on the luminometer 

E2 

Temperature out 

of range

 

 The luminometer is being used outside of the 

specified operating temperature range (see 
section 

10

 The luminometer has been stored in an 

environment which is outside of its specified 
operating temperature range 

 Luminometer damaged or faulty 

E3 

Erratic 

measurement

 

 Luminometer environment unstable or used in 

an area of high electromagnetic noise 

 Luminometer tilted while measurement being 

performed 

 Luminometer lid damaged and allowing light in 

 Protective pocket dirty or severely scratched 

 

Luminometer damaged or faulty 

E4 

User settings 

undefined

 

 The user configurable settings are undefined - 

check and re-set the 

CLOCK

 format, 

User:

Plan:

 and 

Prog:

 data  

E5 

Program, Test 

Plan or User ID 

undefined

 

 Program thresholds not defined, and have been 

reset to the default values (10

 and

 

30

 Test Plan data not defined 

 User ID not defined 

E6 

Calibration self-

checks failed

 

 Power the luminometer off and then back on 

 Luminometer operating environment unstable 

 Protective pocket dirty or severely scratched 

 Lid not fully closed 

 Lid seal damaged 

 Protective pocket damaged 

 

Luminometer damaged or faulty 

E7 

Internal memory 

failure

 

 Batteries are low or loose 

 Luminometer memory damaged or faulty 

E8 

Internal reader 

fault

 

 

Batteries are flat or loose 

 

Luminometer sample reader is damaged or 

faulty 

E9 

Internal error

 

 

Batteries are flat or loose 

 

Luminometer dropped or subjected to shock or 

vibration 

 

Luminometer damaged or faulty 

Содержание EnSURE v2

Страница 1: ...I Ultrasnap SureTrend and dataSURE II are trademarks of Hygiena LLC Operating Precautions and Limitations of Use The following symbol is used in this manual Description CAUTION WARNING IMPORTANT Hygie...

Страница 2: ...P PROGRAM TEST POINTS AND PASS FAIL LIMITS _____________________ 10 3 3 TEST PLANS ______________________________________________________ 11 3 4 ESTABLISHING ATP PASS FAIL LIMITS _____________________...

Страница 3: ...__ 23 7 TROUBLESHOOTING ________________________________________________ 24 7 1 LUMINOMETER BEEPS _______________________________________________ 24 7 2 TROUBLESHOOTING TIPS __________________________...

Страница 4: ...evels of contamination on surfaces water and product samples 1 Remove swab from test tube and collect sample Refer to instructions included with testing device 2 Place swab back in test tube 3 Break S...

Страница 5: ...ased from inside these cells along with any free ATP picked up from the surface by the swab is now available to react with the chemistry in the bulb of the device When the Luciferin Luciferase enzymes...

Страница 6: ...ng Practices GMP or sanitation plan 1 4 Test Devices and Accessories Device Cat No Description UltraSnap US2020 Surface total ATP test device SuperSnap SUS3000 High sensitivity total ATP test suited f...

Страница 7: ...lect and Exit button Up increase button Start accept button Down decrease button Program select button Statistics shortcut button The functions of the keypad are explained in detail in the following s...

Страница 8: ...Cell Voltage Relative Capacity Alkaline 1 5V 1 0 NiMH 1 2V 0 6 NiCD 1 2V 0 5 The unit requires two batteries of size AA LR6 or E91 TIP For best results always use a quality brand of Alkaline battery...

Страница 9: ...ely The unit will then perform an internal 15 second calibration see section 2 7 NOTE If the clock is not set the unit will automatically enter the time and date set up mode see section 3 8 before per...

Страница 10: ...ement reading in Relative Light Units RLU Retested icon identifies the failed test result as having been retested see section 4 2 Retest result icon identifies the Program location or test result as a...

Страница 11: ...and is flat TIP The button can be pressed at any time to access the set up menu options see section 3 When the calibration is complete the luminometer is ready to perform a sample measurement For deta...

Страница 12: ...button Use the and buttons to scroll through the menu list and then press the button to select the required menu option or the button the exit the menu The following 11 menu options are available User...

Страница 13: ...uminometer or set using the SureTrend software If a location name is entered and threshold set in SureTrend and then downloaded to the luminometer the threshold values can t be changed via the luminom...

Страница 14: ...ocations specified in the order in which they would normally be tested Having selected the TEST PLAN menu option use the and buttons to scroll through the list of Test Plans and then press the button...

Страница 15: ...guide included with your system or visit the resources available on www hygiena com 3 5 Viewing Stored Test Results The luminometer can store a maximum of 2000 test results in its internal memory eve...

Страница 16: ...to exit the option 3 8 Setting the Clock Time and Date To set or change the clock time date and format select the CLOCK menu option by pressing the button Scroll to CLOCK and press Then use the and bu...

Страница 17: ...ready to perform readings Quick Boot can be turned off and the luminometer will perform a more in depth system check the next time it is turned on The more in depth system check takes 60 seconds NOTE...

Страница 18: ...able or disable the warning beeps This will toggle the bell icon from enabled to disabled Press the button to accept the settings increase or decrease duration 5 to 60 set duration enable or disable w...

Страница 19: ...corded NOTE The Save Tests mode will be set back to ON when the luminometer is turned off and back on again When Save Tests mode is set to OFF RESULTS NOT SAVED will appear at the bottom of the LCD To...

Страница 20: ...y shows the following details Various keypad options are now available as detailed in each of the following sections Button Action Section Access set up menu options 3 Select different User ID 3 1 Sel...

Страница 21: ...e displayed WARNING Always ensure that the exterior of the testing device is clean and dry before inserting it into the luminometer Never insert anything other than a Hygiena testing device into the l...

Страница 22: ...location was a failure Having selected retest mode signified by the icon next to the Prog perform the sample measurement in the usual way using a new testing device see section 4 1 Once the retest mea...

Страница 23: ...on your computer has been disabled Open My Computer right click on the CD drive and select AutoPlay If this does not start the install program open the CD folder and run the Setup application NOTE Yo...

Страница 24: ...ardware Wizard does not start check the following a You have sufficient permissions to run the wizard b Verify there are no existing hardware problems in device manager c Verify the Universal Plug and...

Страница 25: ...intenance The luminometer does not require any specific routine operator or service engineer maintenance 6 1 Cleaning the Luminometer Case Clean the luminometer case when required using a dry or sligh...

Страница 26: ...ocket using water or a very mild detergent solution ensuring that the pocket is completely dry and clean before putting it back into the luminometer WARNING Do not use solvents or other strong chemica...

Страница 27: ...use the luminometer emits a variety of different beeping sounds Beep Type Possible Causes Short high pitched tone Luminometer turned on or off Sample measurement started Communications established wit...

Страница 28: ...window is dirty Display window is scratched or dented Display or luminometer damaged or faulty Keypad button has no effect when pressed Some buttons only work when selecting particular luminometer fu...

Страница 29: ...dirty or severely scratched Luminometer damaged or faulty E4 User settings undefined The user configurable settings are undefined check and re set the CLOCK format User Plan and Prog data E5 Program...

Страница 30: ...n of any part of the luminometer or the use or attachment of any peripheral not supplied or specified by Hygiena will void this Warranty WARNING Use only the accessories and consumables supplied by Hy...

Страница 31: ...t resolution 1 RLU Measurement time 15 seconds Measurement noise 5 or 5 RLUs User IDs 200 IDs of 16 characters Programmable result thresholds 5001 programs Programmable location names 5001 names of 20...

Страница 32: ...alkaline batteries or rechargeable NiMH NiCD batteries of types specified in section 10 WARNING Do not use batteries with individual cell voltages greater than 1 65V as this will cause permanent damag...

Страница 33: ...1 The luminometer is manufactured under ISO 9001 controls Declaration of Conformity The luminometer has been designed in accordance with and satisfies the requirements of article 11 of the Low Voltage...

Отзывы: