INS0213P Rev. C 

May 2021 

CalCheck for EnSURE

 Touch 

LED Calibration Verification for Hygiena Luminometers 

Cat No.: CAL  

Description/ Intended Use: 

CalCheck is a reusable device. Hygiena recommends incorporating 
CalCheck into a quality control program to verify that the luminometer 
continues to operate correctly and is within specifications. When powered 
on by pressing the button on CalCheck's cap, the device emits a 3-second 
green LED light output. In the off state, CalCheck can be used as a 
negative control to alert users to background instrument light leakage or 
rare sensor calibration errors. Hygiena recommends verification of 
luminometer calibration with CalCheck at a regular interval (e.g., weekly, 
monthly) as specified by user-defined quality procedures. 

Instructions: 

Do not twist or bend CalCheck. 
1. Turn on EnSURE™ Touch 
2. Press the Pink Check Calibration App 
3. Follow steps on screen to initiate negative calibration check. 
4. Follow steps on screen to initiate positive calibration check.  
5. Upon completion, results are automatically recorded in the instrument 
and will be synced to your SureTrend™ Cloud account when connected to 
WiFi. 
 

 

Troubleshooting:  

Higher-than-expected CalCheck RLU readings indicate system calibration 
failure. Physical contamination can cause lower-than-expected CalCheck 
RLU readings. 

 

Clean outside of rod with lint-free cloth containing isopropyl or ethyl 
alcohol. Air dry and repeat reading.  

 

F

or instructions on how to clean EnSURE™ Touch please visit: 

https://help.hygiena.com/kb-doc/how-to-clean-the-instrument/

 

 

If issue is not resolved, contact Hygiena for technical support. 

Incorporating CalCheck into a Quality Control Program: 

CalCheck is designed to be incorporated into a quality control program that 
monitors and tracks performance of Hygiena luminometer(s). For optimum 
recordkeeping, it is recommended that a specific location be programmed in 
SureTrend

™ Cloud data analysis software for measuring CalCheck on 

luminometer.  

Storage & Shelf Life:

  

 

Store at 10 

– 25 °C (50 – 77 °F) in original packaging.  

 

If CalCheck is exposed to extreme temperatures (direct sunlight, heat, 
refrigeration, or frozen), allow device to acclimate to ambient 
temperature (10 

– 25 °C / 50 – 77 °F) for 30 minutes before use. 

 

Take care to not press power button during storage as this will shorten 
battery life. 

 

Replace CalCheck as needed. Refer to expiration date on label. 

 
 
 
 
 
 
 
 
 

 

Battery Replacement:

  

 

When button cap glows red, CalCheck is not ready to perform a 
positive check and may be approaching end of battery life. Wait 2 
minutes and try again. If red light continues to appear, battery should 
be replaced. 

 

If light remains dark upon pressing button, battery should be replaced. 

 

CalCheck is powered by 1-CR1025 coin lithium battery (3V) available 
from any common battery supplier. Battery door is located on 
CalCheck’s black cap. Remove by inserting your fingernail along edge 
of door and gently removing cover. Insert new battery and replace 
door. 

 
Caution: 

Wait 2 minutes between each positive reading. Repeatedly activating 
CalCheck may drain battery prematurely. 

 

Disposal: 

CalCheck contains RoHS-compliant materials and a small battery. With 
battery in place, CalCheck should be discarded as electronic waste. With 
battery removed, CalCheck may be disposed as trash. CAUTION 

– RISK 

OF EXPLOSION IF BATTERY IS REPLACED BY AN INCORRECT TYPE. 
DISPOSE OF USED BATTERIES ACCORDING TO THE INSTRUCTIONS. 
 

Warranty:  

Hygiena warrants CalCheck, when purchased new, to be free from defects 
in materials and workmanship, and will repair or replace, at its discretion, 
any CalCheck device which, used under proper conditions, exhibits such 
defects. Under the terms of this warranty, the product must be returned in 
the original packaging, transportation prepaid, with a copy of the Proof of 
Purchase, to your local distributor. Contact your local distributor to receive 
authorization to return the instrument, and enclose a detailed description of 
the problem. This warranty is provided to the original purchaser for one year 
from the date of purchase. In no event will Hygiena be liable for indirect, 
incidental or consequential damages; the original user’s remedies being 
limited to repa

ir or replacement of CalCheck at the manufacturer’s option. 

Unauthorized modification of any part of CalCheck, including removal of 
labeling will void this Warranty. 

 
Contact Information:  

If more information is required, please visit 

www.hygiena.com

 or contact: 

 

Hygiena 

— Americas 

Phone: 1.805.388.8007 
E-mail

[email protected]

 

Hygiena 

— Internationa

Phone: +44 (0)1923 818821 
E-mail: 

[email protected]

 

 