Configuration and Calibration Guide HD0679 Rev 1.2.0 45
1.4
Symptom: Near Constant Moisture Reading
Possible
explanation
Check
Required result
Action required on
failure
Empty bin or sensor
uncovered
Sensor is covered by
material
100mm minimum
depth of material
Fill the bin
Material stuck in bin
Material is not stuck
above sensor
Smooth flow of
material over the
sensor face when the
gate is open
Look for causes of
erratic flow.
Reposition sensor if
problem continues
Build-up of material
on sensor face
Signs of build-up
such as dried solid
deposit on ceramic
face
Ceramic faceplate
should be kept clean
by the action of
material flow
Check angle of the
ceramic in range of
30° to 60°. If problem
continues reposition
sensor.
Incorrect input
calibration within
control system
Control system input
range
Control system
accepts output range
of sensor
Modify control
system, or
reconfigure sensor
Sensor in alarm
condition – 0mA on
4-20mA range
Moisture content of
material by oven
drying
Must be within
working range of
sensor
Adjust sensor range
and/or calibration
Interference from
mobile phones
Use of mobile
phones close to
sensor
No RF sources
operating near to
sensor
Prevent use within
5m of sensor
Average/Hold switch
has not operated
Apply signal to digital
input
Average moisture
reading should
change
Verify with Hydro-
Com diagnostics
No power to sensor
DC power at junction
box
+15Vdc to +30Vdc
Locate fault in power
supply/ wiring
No sensor output at
control system
Measure sensor
output current at
control system
Varies with moisture
content
Check cabling back
to junction box
No sensor output at
junction box
Measure sensor
output current at
terminals in junction
box
Varies with moisture
content
Check sensor output
configuration
Sensor has shut
down
Disconnect power for
30 seconds and retry
or measure current
drawn from power
supply
Normal operation is
70mA – 150 mA
Check operating
temperature is within
specified range
Internal failure or
incorrect
configuration
Remove sensor,
clean face & check
reading (a) with
ceramic face clean
and (b) with hand
pressed firmly on
ceramic face.
Reading should
change over a
reasonable range
Verify operation with
Hydro-Com
diagnostics
Содержание Hydro-Mix
Страница 4: ...4 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Страница 6: ...6 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Страница 8: ...8 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Страница 10: ...10 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Страница 14: ...Chapter 1 Introduction 14 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Страница 28: ...Chapter 2 Configuration 28 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Страница 40: ...Chapter 3 Sensor Integration and Material Calibration 40 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Страница 42: ...Chapter 4 Best Practices 42 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Страница 48: ...Chapter 5 Sensor Diagnostics 48 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Страница 52: ...Appendix A Document Cross Reference 52 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...