Hydro-View User Guide HD0531 Rev 2.0.0 71
The calibration data points define a mathematical best fit line and it is this line, described using
the variables A, B and C, which defines the calibration. The effect of the Quick Start Rules is to
improve this calibration line if the calibration data does not satisfy the criteria as described in
Appendix C. In such cases the mathematical best fit line is modified. The Quick Start Rules can
be used where the calibration samples taken do not result in a large enough moisture variation to
produce an accurate calibration. When the moisture in the material changes enough to enable
calibration samples to be taken over a greater moisture spread the Quick Start Rules should no
longer be used.
It should be noted that the Quick Start Rules have been designed around a sensor mounted at
the suggested angle. See the individual sensor user manuals for more information.
Hydro-View enables the selection of one of five Quick Start material types:
•
0-2mm Sand
•
0-4mm Sand
•
4-8mm Gravel
•
8-16mm Stone
•
16-22mm Stone
If however, different materials are being measured or the installation is different from the
suggested method, then the Quick Start Rules should be disabled. This is application specific
and should be determined by the engineer commissioning the equipment.
In the graph below, two calibration points have been entered into the table with the Quick Start
Rules enabled. The data does not satisfy the complete criteria and as a result, a warning
message is displayed as shown. The B and C calibration coefficients that describe this line have
been modified.
Figure 75: Quick Start Rules Applied
4 Calibration Procedure
4.1
Equipment Required
To collect calibration samples the following equipment is required:
•
Microwave oven or alternative heat source
•
Weigh scales 2kg accurate to 0.1g
•
Microwave safe bowls
•
Metal spoon
•
Heat proof gloves and eye protection
•
Heat resistant mat, placed on the scales to protect it from the heat of the bowl
Содержание HV04
Страница 4: ...4 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 6: ...6 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 12: ...12 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 14: ...14 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 22: ...Mechanical Installation Chapter 2 22 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 28: ...Electrical Installation Chapter 3 28 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 78: ...Material Calibration Chapter 6 78 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 80: ...Default PIN Codes Appendix A 80 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 82: ...USB Memory Stick file format Appendix B 82 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 84: ...Quick Start Rules Appendix C 84 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 86: ...Frequently Asked Questions Appendix D 86 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 92: ...Document Cross Reference Appendix G 92 Hydro View User Guide HD0531 Rev 2 0 0 ...