Hydro-View User Guide HD0531 Rev 2.0.0 83
Appendix C
Quick Start Rules
1 Quick Start Rules
•
Limiting slopes (B) for any calibration will be a maximum of 2.0 and a minimum of 0.06.
•
One-point calibrations:
•
The calibration slope will be set to be the average of the two known sand calibrations.
•
If the Unscaled value at zero moisture is less than 5 the Unscaled value at zero
moisture will be set to be 5 and a new calibration slope calculated through this point
and the single point entered.
•
If the Unscaled value at zero moisture is greater than 50 the Unscaled value at zero
moisture will be set to be 50 and a new calibration slope calculated through this point
and the single point entered.
•
If the resulting slope is greater than the maximum or less than the minimum calibration
slopes then no calibration will be carried out and the user will be informed of this.
•
More than 1 point calibration – spread in points: Moisture < 1% or Unscaled < 2
•
A single point calibration will be carried out.
•
More than 1 point calibration – spread in points: Moisture < 3% or Unscaled < 6
•
If the calculated slope is greater than the selected materials quick start calibration
slope set the calculated slope to the quick start calibration slope - If the calculated
slope is less than the selected quick start calibration slope set the calculated slope to
the quick start calibration slope – otherwise leave the slope as it is. (Recalculate
intercept value from average from all points)
•
If the Unscaled value at zero moisture is less than 5 the Unscaled value at zero
moisture will be set to be 5 and a new calibration slope calculated through this point
and the average of the points entered.
•
If the Unscaled value at zero moisture is greater than 50 the Unscaled value at zero
moisture will be set to be 50 and a new calibration slope calculated through this point
and the average of the points entered.
•
If the resulting slope is greater than the maximum or less than the minimum calibration
slopes then no calibration will be carried out and the user will be informed of this.
•
More than 1 point calibration – spread in points: Moisture > 3% and Unscaled > 6
•
Calibration slope is calculated and the user is warned if:
•
If the Unscaled value at zero moisture is less than 5.
•
If the Unscaled value at zero moisture is greater than 50.
•
If the resulting slope is greater than the maximum or less than the minimum
calibration.
Содержание HV04
Страница 4: ...4 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 6: ...6 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 12: ...12 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 14: ...14 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 22: ...Mechanical Installation Chapter 2 22 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 28: ...Electrical Installation Chapter 3 28 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 78: ...Material Calibration Chapter 6 78 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 80: ...Default PIN Codes Appendix A 80 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 82: ...USB Memory Stick file format Appendix B 82 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 84: ...Quick Start Rules Appendix C 84 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 86: ...Frequently Asked Questions Appendix D 86 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 92: ...Document Cross Reference Appendix G 92 Hydro View User Guide HD0531 Rev 2 0 0 ...