Control Panel Basic (SIMATIC PANEL)
FAM 25/45/60/75/95
en(us)
Page 66/160
BEWA FAM25-95 3402054f en(us) 2013-06-13
2013-06-13
T (Temperature)
MENU1 -> T
F1
F2
F3
F4
START =
Start
FAM
Display in operating status "Ready for operation" and with the
error message "Water content
≥
90%, measurement with CS
not possible".
STOP =
Stop
FAM
is displayed only in the operating status "Starting phase" and
"Automatic operation"
RESET
= Reset error messages
Is displayed only in the presence of faults
MENU1
= Switch to MENU1
Display :
Hysteresis:
= Current hysteresis setting of the heater.
The hysteresis range is adjustable from 3 to 15°C.
Factory setting: 5%
Temperature Actual:
= The heater temperature nominal value can be set
between 0° and 99°C. For details, see page 62.
The safety thermostat responds at 80°C.
Factory setting: 55%.
Temperature Actual:
= Current temperature in the vacuum column.
The displayed temperature is measured by a sensor in the sump of the vacuum
column.
The temperature setting can be changed during automatic operation.
START
MENU1
T
Heater: Hysteresis:
5°C
Temperature: Nominal: 55°C Actual:
Ready for operation. Please press Start
STOP
RESET