Maintenance and Servicing Manual
FAM 200 FluidAqua Mobil
5
Description of the Inspection and Maintenance Work
The pictures and explanations in the descriptions below are examples. They don't show all product
variants.
The unit is to be shut off when performing work, except as specifically indicated otherwise:
Steps:
Shut down automatic mode of the FAM via the display. Wait ca. 1 minute until
normal pressure is restored in the vacuum column (ambient pressure).
The FAM can be switched off as soon as the pressure in the vacuum column has
dropped to normal (ambient) pressure.
Secure the unit from being inadvertently switched back on.
5.1
Function Test of the Fault LED (40)
Steps:
Switch the FAM ON at the main switch.
A lamp test is automatically conducted when switching the unit on,
during which the fault LED lights up briefly.
5.2
Changing the Air Filter (19)
You need:
•
Air filter (for stock no. see
Replacement Parts List
)
Steps:
The air filter has to be changed every six months.
The filter has to be changed more frequently if the FAM is
operated in an extremely dusty or humid environment.
Unscrew the air filter by hand
Screw on a new air filter by hand
(O ring not needed)
HYDAC Filtertechnik GmbH
en
Page 16
IuW FAM200 3373093a en.doc