VII. Operation Instructions

Press the ON/OFF button; turn the measure range switch to choose the

correct testing parameters and range.

1. Measuring AC voltage
Turn the rotary switch to 500V in U1 region, and connect the measured

voltage into jack U1, you can measure. If the measured value is less than 200V,
you can directly turn the switch to 200V in U1 region to obtain better
measurement accuracy.

The two channels U1, U2 are provided with the same voltage test function,

so you can make the measurement through channel U2. Detailed methods
please refer to the operation of U1.

2. Measuring AC current
Turn the rotary switch to 10A in I1 region, and connect the down-lead of the

clamp-current transformer into jack I1, then clamp the measured circuit with
clamp jaw, you can measure. If the measured current is less than 2A, you can
directly turn the switch to 2A in I1 region to obtain better measurement
accuracy.

The two jacks I1, I2 are provided with the same current test function, so

you can make the measurement through jack U2. Detailed methods please
refer to the operation of I1.

3. Measuring phase angle between two voltages
When measuring the phase angle that U2 lags U1, turn the rotary switch to

parameter U1U2.Under measurement, you can turn the rotary switch clockwise
to different measure ranges in U1 region to measure input voltage U1, or turn
the rotary switch anticlockwise to different measure ranges in U2 region to
measure input voltage U2.

Note: When measuring phase, the marks U1, U2 around voltage input

jacks and the red "*" mark on clamp-current transformer are phases with the
same name end.

Содержание SMG2000B

Страница 1: ...al You can find the changes in the appendix Sorry for the inconvenience If you have further questions welcome to contact with our service department The input output terminals and the test column may...

Страница 2: ...r relevant subassembly to damage In order to avoid possible danger this product can only be used within the prescribed scope Only qualified technician can carry out maintenance or repair work To avoid...

Страница 3: ...uire information about the ratings Do Not Operate without Covers Do not operate this product when covers or panels removed Use Proper Fuse Only use the fuse with type and rating specified for the prod...

Страница 4: ...Security Terms Warning indicates that death or severe personal injury may result if proper precautions are not taken Caution indicates that property damage may result if proper precautions are not tak...

Страница 5: ...s I Overview 6 II Intrinsic Error 6 III Operating Error 7 IV Other Technical Specifications 8 V Product Structure 9 VI Notes 9 VII Operation Instructions 10 VIII Adjust the display 12 IX Battery repla...

Страница 6: ...ed by clamp current transformer When measuring the phase between U1 and U2 the two input circuit loops are completely isolated so it can completely avoid short circuit and burning down the meter cause...

Страница 7: ...2 AC current see Table 2 Table 2 Error of AC current Measure range Resolution Limits of intrinsic error 200mA 0 1mA 1 0 RD 2 2A 1mA 10A 10mA 3 Phase U U U I I I see Table 3 Table 3 Error of Power fre...

Страница 8: ...ded 2 Limits of Rated Operating Error In item 1 Rated Operating Conditions the limits of rated operating error should not more than twice limits of intrinsic error IV Other Technical Specifications 1...

Страница 9: ...ce VI Notes 1 Withstand Voltage Parts between voltage input terminal and meter shell and parts between the iron core and handle of the clamp current transformer and secondary windings can withstand 10...

Страница 10: ...into jack I1 then clamp the measured circuit with clamp jaw you can measure If the measured current is less than 2A you can directly turn the switch to 2A in I1 region to obtain better measurement ac...

Страница 11: ...the voltage into U2 input the current into I1 by using measuring clamp I1 and turn the rotary switch to parameter I1U2 to measure the angle that the voltage lags the current Under measurement in the s...

Страница 12: ...nto jack I2 with measuring clamp If the phase value ranges from 0 90 then the measured load is inductive If the phase value ranges from 270 360 then the measured load is capacitive VIII Adjust the dis...

Страница 13: ...13...

Отзывы: