function options: Test, Stop, Save or Disp; press the key <↑>
,
<↓> to modify
the phase of current capacitor under test (ab_bc_ca), while the single-phase
capacitor is invalid. When modified to the required test items, press the
function button Test to enter the capacitance test state, as shown in the left-top
figure. The current measurement data can be temporarily stored by pressing
the function button Save. The capacitance values of Delta or Y-shaped
capacitors can be displayed by pressing the button Disp. When ab→bc→ca
three-phase cyclic test is carried out and temporarily stored, the
phase-separated capacitance value, the compatible capacitance value and the
rated capacity of the tested capacitor can be obtained as shown in the right-top
figure. Press the button Stop to cut off the output of the test power, and the
test is over.
In the Testing Capacitance interface, the current test results can be saved in
local memory by pressing the key < Save> , and the current test parameters
and test results can be printed by pressing the key <print> .
In the Testing Capacitance interface, press the < Back > key to cut off the
output of the test power, and return to the Test Options interface.
6.2.2 Inductance Test
In the Main Menu interface, press <SEL> or the key <↑> and <↓> to select the
Test, and then press the key <OK> to enter the Test Options interface. When
performing the inductance test, you need to set the test parameters first.
Select 1.Test SET to enter the Trial Parameter Setting interface, and the
operation is same with that of capacitance test function.
Note: in the trial parameter setting the connection mode is invalid for the
inductance test parameter.