13
Press the select key move cursor - to the year, month, day, time
and distance, press the + or - key to choose the specific numerical,
according to the confirmation key, and return to the start page;
6. In figure 2 page, press the select key to move the cursor)
Settings, press the ok button to enter the custom set child pages
(figure 6);
Figure 6 a custom Settings page
7. In figure 6 pages, press the select key move the cursor to the
corresponding option, then press the + or - key parameters can be
set.Among them:
Let stand for 15 min, the default value range of 1 ~ 15 min
(incremental Δ = 1 min);
The time interval between the default value of 5 min, range 1 ~ 10
min (incremental Δ = 1 min);
Mixing time default 10 s, the range of 5 ~ 90 s (incremental Δ = 5 s);
Stop up the default 60 kV voltage, range of 10 ~ 80 kV (incremental
Δ = 10 kV).When the instrument booster to stop rising voltage Time