1 detection switch
2 power switch
3 power outlet
4 ac fuse (0.5A)
5 dc fuse (5A)
6 the host
Seven of the needle valve
8 handheld probe
9 LCD
10 buttons
11 ionization chamber observation window
12 gas electric tube

IV. Operation procedure and method

1. Plug in the power supply and turn on the "power switch", then the motor's
starting sound can be heard.
Turn on the "detection switch". If you can see a faint dark purple excitation light
(" initiation ") in the window above the probe, and the base value of the lower
side of the LCD screen is close to the value marked on the nozzle of the
instrument, the measurement can start.(after starting up, the large value in the
middle of the display screen is the concentration of SF6 gas detected in real
time, so it is unnecessary to check the table again. The small value below is
the base value, which is only used as a reference for the stability of the
instrument.)
3. If not, you can plug the inlet of the probe (needle valve) with your finger to
improve the vacuum of the vacuum system.If the probe is blocked within 10
seconds, wait 10 minutes and then block the probe again. The time of blocking
the probe shall not exceed 10 seconds, otherwise the relevant electronic
devices will be damaged.
4. The instrument will start up for 60 seconds and set the zero point by
self-calibration. If the zero point is not reached within 60 seconds, the probe
will be held in the air and press "cancel" to set the zero point.
5. After 10 minutes, after the instrument reaches a certain vacuum degree and
the base value is basically stable, move the probe to the place under

Содержание HTJL-H

Страница 1: ...al You can find the changes in the appendix Sorry for the inconvenience If you have further questions welcome to contact with our service department The input output terminals and the test column may...

Страница 2: ...r relevant subassembly to damage In order to avoid possible danger this product can only be used within the prescribed scope Only qualified technician can carry out maintenance or repair work To avoid...

Страница 3: ...uire information about the ratings Do Not Operate without Covers Do not operate this product when covers or panels removed Use Proper Fuse Only use the fuse with type and rating specified for the prod...

Страница 4: ...Security Terms Warning indicates that death or severe personal injury may result if proper precautions are not taken Caution indicates that property damage may result if proper precautions are not tak...

Страница 5: ...5 Contents Warning 6 I Product introduction 6 II Technical data 7 III Product structure 7 IV Operation procedure and method 8 V Adjustment method 9 VI Announcements 11...

Страница 6: ...The concentration degree of sulfur hexafluoride can be displayed by a detection of the degree of ionization This instrument is a small vacuum pump with its searching unit and main engine separated Tho...

Страница 7: ...ppm Volume ratio Pumping speed of vacuum pump 0 5 litre second Length of the vacuum tube 4 m Continual working time not more than 3 hours Working condition alternating current power supply 220V 50Hz E...

Страница 8: ...e concentration of SF6 gas detected in real time so it is unnecessary to check the table again The small value below is the base value which is only used as a reference for the stability of the instru...

Страница 9: ...m value using The key changes When the volume ratio is greater than the alarm point the device will emit a beeping alarm V Adjustment method 1 Making up different concentration degree sulfur hexafluor...

Страница 10: ...to be demarcated Press OK and then demarcate the first point Input OPPM and then press OK Input 1010 which is the first point s collecting number Press OK and then QUIT to enter the demarcation interf...

Страница 11: ...ppropriate amount The three adjustable potentiometer inside the searching unit has been set for its output voltage and signal output value before selling In generall it doesn t need to be adjusted aga...

Отзывы: