HFR-8000 Operator
’s Manual for Remote Tracing
31
Pattern Calibration
•
Purpose
Detecting the pattern-reading position, axis, and calibrating the size.
•
Procedure
①
Load the pattern holder with the standard pattern attached
②
Press MENU button and select the
“CALIB >> PATTERN” menu
Refer to the section 6.1 for menu operation
③
Press the BOTH button two times to trigger the automatic calibration
It traces the standard pattern twice
④
If it result in any error, check if the standard pattern is loaded in correct position
and try again
If it fails again, contact proper service engineer for maintenance
•
Preparation for the PATTERN calibration
•
PATTERN calibration menu
Pattern Holder
Standard Pattern
B: Execute
[ CALIB ] PATTERN