background image

 

O

Operator’s Manual 

 

HFR-8000 

For Remote Tracing 

 
 
 
 

 

 
 
 

 

 

Содержание HFR-8000

Страница 1: ...Operator s Manual HFR 8000 For Remote Tracing...

Страница 2: ...ate at the time of publication HUVITZ assumes no responsibility however for possible errors or omissions or for any consequences resulting from the use of the information contained herein HUVITZ reser...

Страница 3: ...5 2 Safety Information 6 2 1 Introduction 6 2 2 Safety Symbols 7 2 3 Environment factors 8 2 4 Safety Precautions 11 3 System Configuration 15 3 1 Frame Reader 15 3 2 Standard Accessory 16 4 Installa...

Страница 4: ...7 1 Interface Specification 24 7 2 Connecting to a DCS host 25 7 3 DCS Options 26 7 4 Using the Barcode Interface for Inputting Job Number 27 8 System Maintenance 29 8 1 Automatic Calibration of Fram...

Страница 5: ...th e stylus position into the frame groove especially when the groove is lopsided SEMI AUTO demo lens tracing mode is additionally provided for recognizing and processing concave shape Barcode scanne...

Страница 6: ...r s Manual for Remote Tracing 5 in an environment with continuous operation over a long period of time 1 3 Classification Protection against electric shock Class I earthed Installation Category II Pol...

Страница 7: ...the user operator It is for this reason that the following safety notices have been placed appropriately within the text of this manual to highlight safety related information or information requirin...

Страница 8: ...afety ground is fastened to the chassis Protective earth connected to conductive parts of Class I equipment for safety purposes Disposal of your old appliance 1 When this crossed out wheeled bin symbo...

Страница 9: ...use only Where the machine is exposed to direct sunlight Where there are big changes in temperature Optimal temperature range for normal operation is from 10 C to 40 C Humidity 30 80 Where there is a...

Страница 10: ...ill cause severe damage on the machine Pull out the power cord with holding the plug not the cord To avoid risk of electric shock this equipment must only be connected to a supply mains with protectiv...

Страница 11: ...ge of 10 95 with non condensing An atmospheric pressure range of 700 1060hpa 4 Temporary Overvoltage Short term temporary overvoltage 1440V Long term temporary overvoltage 490V Please avoid where the...

Страница 12: ...ar chemicals Do not put or use this device in humid rooms Humidity should be maintained between 50 and 80 for normal operation Do not expose the device to water splashes dripping water or sprayed wate...

Страница 13: ...authorized to install use operate and maintain this equipment Keep accompanied manual s or technical materials such as the User s Manual and Service Manual in a place easily accessible at all times fo...

Страница 14: ...mble the product without the proper training It may cause the electric shock serious hurt during the operation or cause the malfunction of the product The product should be properly installed and main...

Страница 15: ...sure the locking device and the screw for the cover are properly tightened In case of packaging the equipment use the recommended packaging material and method in order to prevent the damage during th...

Страница 16: ...System Configuration 3 1 Frame Reader Front View Rear View Protection Door Control Panel Work Space for Stylus Display Window Gripper Gripping Bar Stylus Main Switch Power Inlet Fuse Holder Interface...

Страница 17: ...16 HFR 8000 Operator s Manual for Remote Tracing 3 2 Standard Accessory Pattern Holder Standard Pattern Standard Frame...

Страница 18: ...e the shock absorbing material from the workspace of Stylus Plug the Power Cable into the power inlet at the back of the product and turn on the power to check the initial operation If the initial ope...

Страница 19: ...cing Mode LO CURV 2D Tracing HI CURV 3D Tracing When the Pattern Holder is loaded in the Frame Reader it automatically recognizes the pattern tracing mode When tracing a demo lens of higher curve usin...

Страница 20: ...n one of the tracing buttons for 2 seconds will executes the SEMI AUTO tracing mode which is to allow recognizing concave shape while tracing Otherwise it doesn t recognize concave shape and reform it...

Страница 21: ...gripping bar Select the frame material by the material selection button Press one of the tracing buttons BOTH for both sides R for right or L for left Pattern Tracing Put a pattern or a blocked demo l...

Страница 22: ...reverses the selection FRAME button selecting a configuration parameter in a menu category Holding down the button for 2 seconds reverse the selection BOTH button executing a selected parameter for c...

Страница 23: ...the port EDGER PC NONE PROTO 1 Data communication protocol for EDGER 1 port HUVITZ DCS DCS_Z BPS 1 Baud rate for EDGER 1 port 9600 14400 19200 57600 115200 38400 EDGER 2 Target device for EDGER 2 port...

Страница 24: ...g mode YES NO MINI FR Low speed tracing option for frames with minimal vertical size YES NO FPD OFF FPD offset 2 99 2 99 JOB NO Job number reset option Turning on the option keep the last job number e...

Страница 25: ...specification of the communication port of the Frame Reader for interfacing PC follows the information below Baud Rate 115200 57600 38400 19200 14400 or 9600 Data Bit 8 Stop Bit 1 Parity None Flow con...

Страница 26: ...ing a lab management software or a host program on it make it sure to turn the instrument off Procedure Connect PC to EDGER1 port with a 9 pin DSUB crossed serial cable Connect a barcode reader to the...

Страница 27: ...1 for menu operation Set at least one of EDGER 1 and EDGER 2 port like bellows EDGER 1 2 PC PROTO 1 2 DCS or DCS_Z Change the menu category to DCS DCS menu category is activated only if one of the edg...

Страница 28: ...ngle side is traced YES NO FRAME Default frame material for a job for DCS output METAL Metal PL HARD Hard Plastic PL SOFT Soft Plastic UNDEFIN Undefined OFF Do not output frame material info METAL PL...

Страница 29: ...ps and prompts to scan a barcode Scan a barcode and it ll display the number for a second on the screen Frame tracing is automatically started after the job number input The job number is transmitted...

Страница 30: ...for frame reader consists of three parts Stroke Frame Pattern In case of any change happened in the stylus do the full calibration in the following sequence Stroke calibration Frame calibration Patte...

Страница 31: ...n menu Calibration results are stored in its permanent memory automatically right after the completion of each calibration Be sure to check the related information before executing the calibration wor...

Страница 32: ...B PATTERN menu Refer to the section 6 1 for menu operation Press the BOTH button two times to trigger the automatic calibration It traces the standard pattern twice If it result in any error check if...

Страница 33: ...ess MENU button and select the CALIB STROKE menu Refer to the section 6 1 for menu operation Press the BOTH button two times to trigger the automatic calibration If it result in any error check the mo...

Страница 34: ...t the power cord Since the exterior of the product is easily scratched be sure to use soft cleaning cloth When cleaning the product avoid spraying cleansing agent directly onto the product This may re...

Страница 35: ...n for Frame Reader Calibration 1 Pattern Holder 1 Fuse 3 15A 2 Power Cable Frame Reader 1 Option Barcode Scanner 1 9 pin D Sub Serial Cable Crossed 1 Lens Adapter 3 Specification Frame Reader Dimensio...

Страница 36: ...HFR 8000 Operator s Manual for Remote Tracing 35 System Dimension...