Getting started instructions for Access DH2 Prober 

21-2627 Rev A Version 1.4 

System Power Up

 

1.

 

Verify the Front Panel Power switch is off (rotary switch located top right) 

2.

 

Verify the Front Panel Emergency Shutdown switch is not engaged (located under the 
Front Panel Power switch). 

3.

 

Verify that the PDU breaker is OFF and the toggle switch is in the Remote position 
(located in the top position of the equipment rack, inside the front doors). 

4.

 

Verify that the Electronic Chassis power switch is off (located in the middle position of the 
equipment rack, inside the front doors). 

5.

 

Bring the power cord from inside the unit out one of the access holes (Remove and 
replace rear and side panels as needed). 

 

NOTE: The DH2 has a US 20A power cord installed that requires a 20A outlet. This is due to the 
capabilities of the internal PDU and ETL / CE safety requirements. If no other equipment is 
added the DH2 system 15A power cord will be sufficient. There is a 15A power cord provided in 
the accessory box. 

 

6.

 

Plug the power cord into an appropriate power source. 

7.

 

Turn on the PDU Breaker. 

8.

 

Turn on the Front Panel Power switch. 

9.

 

Turn on the monitor. 

10.

 

Turn on the PC (located inside the front doors). 

11.

 

Turn on the Electronic Chassis (located inside the front doors). 

 

Camera Lenses (Access DH2 with Blue DFK Cameras)

 

The DH2 has 12mm camera lenses installed in the cameras. These lenses allow for auto focus of 
all four slot levels (Top, Middle, Bottom and Base).  
The accessory kit contains 16mm lenses. The 16mm lenses can provide improved image 
resolution that is useful when working with very fine pitched devices (0.5mm and smaller) in the 
lower Prober slots. The 16mm lenses do have a limited focus range and can auto-focus only on 
one or two slots without having to manually adjust them. Because of this limitation it is 
recommended that the 16mm lenses being used with the Middle, Bottom and Base slots only.  
To change the lenses. Remove the two screws that mount the ring light brackets. Replace the 
lenses. Reinstall the ring lights. In Workstation, Go to Tools->Options->Hardware and change the 
DH2 Lens setting from 12mm to 16mm. Be sure to do a camera offset when changing or focusing 
lenses. 

Access DH2 with Pixelink Cameras

 

The current DH2 configuration uses high quality auto-focus cameras from Pixelink. When 
configuring your Huntron Workstation set up (Tools menu/Options/Hardware tab) be sure to 
select your Frame Grabber type as 

Pixelink

.  

 