5.14 Buzzer Replacement
Disconnect connector CON1. Cut wires from existing buzzer mounted on base.
Fit replacement buzzer onto a clear base area close to the front of the pump
ensuring that the lead is able to reach and fit onto
CON1
.
Use Cyanoacrylate 495 adhesive to bond the sounder’s copper side to the plastic
case.
Ensure the sounder is firmly attached by applying pressure for 10 seconds.
Note: Sounder clarity and volume are dependant on the quality of this bonding
so ensure sounder output is checked by forcing the pump into a fault condition
in order to check sounder output.
5.15 Hang Strap Assembly
Place the unit upside down on a flat, clean surface. Unscrew and remove two
retaining screws from the hang strap plates anchoring the assembly to the base
of the pump unit. Lift the assembly away from the unit and replace it.
247901
5-7
Issue 1
March 1998
Содержание Flowtron AC550
Страница 1: ...247901 Issue 1 Mar 1998 AC550 Flowtron Excel system Service Manual Huntleigh HEALTHCARE...
Страница 2: ...247901 Issue 1 Mar 1998...
Страница 4: ...247901 ii Issue 1 March 1998...
Страница 6: ...247901 iv Issue 1 March 1998...
Страница 12: ...1 4 Pneumatic Schematic 247901 1 4 Issue 1 March 1998...
Страница 13: ...1 5 Valve Sequence 247901 1 5 Issue 1 March 1998...
Страница 14: ...1 6 Electrical Wiring Schematic 247901 1 6 Issue 1 March 1998...
Страница 15: ...1 7 General Assembly 247901 1 7 Issue 1 March 1998...
Страница 16: ...247901 1 8 Issue 1 March 1998...
Страница 18: ...247901 2 2 Issue 1 March 1998...
Страница 30: ...247901 4 8 Issue 1 March 1998...
Страница 32: ...5 4 Replacing Main PCB or Front Display Panel 247901 5 2 Issue 1 March 1998...
Страница 33: ...5 5 Replacing the Motor Gearbox Assembly 247071 247901 5 3 Issue 1 March 1998...
Страница 35: ...5 10 Pressure Control Assembly 247901 5 5 Issue 1 March 1998...
Страница 38: ...247901 5 8 Issue 1 March 1998...
Страница 40: ...247901 6 2 Issue 1 March 1998...