SpectraTrend HT User’s Guide 3.2
58
PRODUCT SETUP, (TAB) PAGE 1: DISPLAY MODE
The available data display modes are
ABSOLUTE
and
DIFFERENCE
.
ABSOLUTE
provides
values as measured by the sensor. These values are relative to the standard white tile
supplied with the system, which in turn is traceable to measurements made by N.I.S.T.
of similar material.
DIFFERENCE
provides values calculated based on the difference
between a product standard’s values and the product sample’s values.
PRODUCT SETUP, (TAB) PAGE 1: ILLUMINANT/OBSERVER
Choose the
ILLUMINANT
and
OBSERVER
that you would like to use when displaying
color measurement values on the front panel of the sensor.
Figure 65. Selection of Illuminant
Product Standard, Tolerances and Corrections
PRODUCT SETUP, (TAB) PAGE 2: PRODUCT STANDARD VALUES
Select the
PAGE 2 TAB
to enter product target values for the color scale selected.
Choose
AD HOC
when you wish to use the first measurement value as the standard.
Choose
NUMERIC
when an actual product specimen does not exist for measurement,
but the target color values are known from previous measurements.
Choose
PHYSICAL
when an actual product specimen is available that represents the
target color to which samples will be compared on a regular basis. A physical standard is
measured and stored in the product setup for as long as desired.
Содержание SpectraTrend HT
Страница 6: ...SpectraTrend HT User s Guide 3 2 6...
Страница 21: ...SpectraTrend HT User s Guide 3 2 21...
Страница 22: ...SpectraTrend HT User s Guide 3 2 22...
Страница 23: ...SpectraTrend HT User s Guide 3 2 23...
Страница 24: ...SpectraTrend HT User s Guide 3 2 24...
Страница 25: ...SpectraTrend HT User s Guide 3 2 25...
Страница 26: ...SpectraTrend HT User s Guide 3 2 26...
Страница 27: ...SpectraTrend HT User s Guide 3 2 27...
Страница 28: ...SpectraTrend HT User s Guide 3 2 28...
Страница 76: ...SpectraTrend HT User s Guide 3 2 76...
Страница 86: ...SpectraTrend HT User s Guide 3 2 86...
Страница 92: ...SpectraTrend HT User s Guide 3 2 92...