Aeros Supplemental Manual for EasyMatch QC ver. 2.0
22
Repeatability Test
The repeatability test is initiated by standardizing and leaving the white tile in view. Each white tile
reading is reported along with a pass/fail evaluation.
Remove all samples from the instrument and attach the standardization box. The sensor will move to the top.
Figure 24. Standardize
Attach the standardization box to the sensor. Then attach the black glass to the standardization box. Press
NEXT
to continue.
Figure 25. Read the Black Glass
Remove the black glass and attach the white tile. Press
NEXT
to continue.
Figure 26. Read White Tile
Leave the white tile at the port press
NEXT
to continue.