Note:
For ease of use, power-up the Hummingbird ICP Control Module and the patient monitor
before beginning the procedure to allow the transducer within the Patient Cable Module
to rise to an equilibrium temperature.
During setup, if repeating the LED and audible chime / beep functional check is desired,
disconnect and reconnect the Monitor Cable from the patient monitor. This action will result
in the functional check being repeated and the system going to ‘Zero Wait Mode’. If you do
not see the expected results on the Hummingbird ICP Control Module or the patient monitor,
check all connections. If you still do not see the expected results, do NOT use the
Hummingbird ICP Control Module. Contact IRRAS USA, Inc. for assistance.
c. Zero patient monitor
. Use the standard hospital protocol to “Zero” the patient monitor.
Remove the catheter or HICP200 Plug (whichever is attached) on the
Hummingbird ICP Control Module Patient Cable Module prior to zeroing
patient monitor.
The Patient Cable Module must communicate with
atmospheric pressure when zeroing the patient monitor. Failure to remove the
catheter and/or HICP200 Plug from the Patient Cable Module prior to re-zeroing
may result in inaccuracy.
d.
Press ‘Zero Monitor’ button on the ICP Control Module.
Note:
Indicator:
The Zero patient monitor indicator light
on the main panel of the ICP Control
Module will go from flashing blue to a
solid blue.