35
Calibrating your instrumentation to the appropriate Input requires a
separate device, which can provide precise and specific loads or dis-
placement, and, which has been certified to be accurate. The calibra-
tion process involves plugging the instrumentation into the HM-5320
or 5330 while placing the instrumentation into the certified calibra-
tion device, which provides a specific set load or displacement.
Once this has been done, click the Calibrate button next to the
appropriate Input
(1)
.
A pop-up window will appear requiring you to enter the password,
which is
22234
.
1