Operation

Control Panel: 

The controls are situated on the front, right hand side of the machine base.

Preset Travel and Speed: 

The shaker has a throw of 8 ± 0.04” and operates at 175 ± 2 cycles per 

minute.

Method:

Prepare the sample to be tested according to the governing agency 

standard(s). Place the clear plastic cylinder containing the test sample and 

rubber stopper in position on the slide assembly of the shaker. The end with 

the spring loaded holder fits over the cork (rubber stopper) in the open end 

of the clear cylinder.
Select the time period appropriate for the test, either 45 seconds or 10 

minutes using the center toggle switch. Then press the start push-button 

switch to begin the test. The test will end at the end of the elapsed time 

selected, or by the operator pressing the pushbutton stop switch any time 

during the test.

Maintenance

First unplug the machine from its power receptacle or source. Normal routine 

maintenance consists of lubrication at several key locations. In addition, worn 

components should be suspected if there is any unusual sound or action of the 

machine.
Lubricate as follows:

A.   The slide assembly should be lubricated once a month with 

approximately three (3) drops of household oil on each end of the main  

slide bar.

B.   Two (2) grease fittings should be greased once a year with automotive 

 

type grease. These are accessible from under the machine through the  

 

fly-wheel.

C.   The cam action guide, H-4374.18 and the two outside slide bars, 

 

H-4374.27 can be greased by smearing a film of grease on each one.

Accessories

In addition to the Sand Equivalent Shaker, Humboldt Mfg. Co. also has available 

a complete Sand Equivalent Test Set that includes four graduated cylinders, 

siphon assembly, weighted foot assembly, wide mouth funnel, 3 oz. measuring 

can and standard stock solution. Consult the Humboldt catalog for this and for 

ordering accessories separately.